Plume generated by electrosurgical techniques is a health hazard to patients and dermatologists.
To compare the particle concentration generated by various energy devices used in dermatologic surgery.
Five surgical techniques were tested on human tissue samples in a closed chamber. A particle counter, positioned at a fixed point 20 cm away from the sample, recorded the concentrations of aerosolized particles generated over 7 particle sizes (0.3, 0.5, 0.7, 1, 2.5, 5, and 10 μm).
Monopolar electrocoagulation created the greatest concentration of particles followed by electrocautery, electrodesiccation, electrofulguration, and bipolar electrocoagulation. Bipolar electrocoagulation created 80 times fewer 0.3 μm particles and 98 times fewer 0.5 μm particles than monopolar electrocoagulation. Across all electrosurgical techniques, the greatest concentrations of particles generated were of the 0.3 and 0.5 μm particle size.
Bipolar electrocoagulation created the lowest concentration of particulate matter. Given the noxious and hazardous nature of surgical plume, the bipolar forceps offer surgeons a safer method of performing electrical surgery for both the surgical staff and the patient.
电外科技术产生的烟雾对患者和皮肤科医生的健康构成危害。
比较皮肤科手术中使用的各种能量设备产生的颗粒浓度。
在封闭室中对人体组织样本进行了五种手术技术测试。位于距样本 20 厘米固定点的粒子计数器记录了在 7 个粒子大小(0.3、0.5、0.7、1、2.5、5 和 10μm)上生成的气溶胶化颗粒的浓度。
单极电凝产生的颗粒浓度最大,其次是电灼、电干燥、电灼和双极电凝。与单极电凝相比,双极电凝产生的 0.3μm 颗粒少 80 倍,0.5μm 颗粒少 98 倍。在所有电外科技术中,生成的颗粒浓度最大的是 0.3μm 和 0.5μm 粒径的颗粒。
双极电凝产生的颗粒物质浓度最低。鉴于手术烟雾的有害和危险性质,双极镊子为外科医生提供了一种更安全的方法来为手术人员和患者进行电外科手术。