Routine sample transportation of centrifuged barrier tubes impacts certain chemistry analytes.

作者信息

Wang Jingcai, Pyle-Eilola Amy L

机构信息

Department of Pathology and Laboratory Medicine, Nationwide Children's Hospital, Columbus, OH, United States.

Department of Pathology and Laboratory Medicine, Nationwide Children's Hospital, Columbus, OH, United States; Department of Pathology, The Ohio State University Medical Center, Columbus, OH, United States.

出版信息

Clin Biochem. 2023 May;115:77-80. doi: 10.1016/j.clinbiochem.2022.08.011. Epub 2022 Sep 2.

Abstract

The preanalytical phase of testing, including sample transportation, is a common source of error in laboratory testing. Previous studies have shown inversion of centrifuged plasma separator tubes (PST) results in elevation in certain analytes. However, it remains unclear if routine transportation practices, without full inversion, can have a similar impact. This study used 12 residual samples submitted in PST and 4 PST tube samples collected from healthy donors. All samples were measured at baseline. Analytes measured were white blood cells (WBCs), red blood cells (RBCs), platelets (PLTs) using the Sysmex XN-1000 and potassium (K), aspartate aminotransferase (AST), lactate dehydrogenase (LDH), sodium (NA), and alanine aminotransferase (ALT) using the Ortho Vitros 4600. The samples were split into two groups: 1 group was placed vertically in racks and the other group were laid horizontally in a messenger bag-style soft-sided cooler used by lab couriers. A cold pack was placed in the bag, equidistant from both groups of samples. To mimic courier transportation, the bag was carried for 30 min, returned to the laboratory and immediately analyzed. Subsequently, the samples were re-centrifuged and analyzed again. Statistics were performed using GraphPad Prism. LDH increased following transportation for vertical and horizontal samples and remained elevated despite re-centrifugation. K was only increased in the horizontal samples following re-centrifugation. Notably, AST as well as WBCs, RBCs, and PLTs all increased following transportation, but dropped to baseline concentrations following re-centrifugation. Centrifuged PSTs should be kept in a vertical position during transportation by courier. Re-centrifugation of plasma gel tubes after transportation may be necessary for certain chemistry tests.

摘要

检测的分析前阶段,包括样本运输,是实验室检测中常见的误差来源。先前的研究表明,离心后的血浆分离管(PST)颠倒放置会导致某些分析物水平升高。然而,尚不清楚常规运输操作(未完全颠倒)是否会产生类似影响。本研究使用了12份提交的PST残留样本以及从健康供体采集的4份PST管样本。所有样本均在基线时进行测量。使用Sysmex XN - 1000检测的分析物为白细胞(WBC)、红细胞(RBC)、血小板(PLT),使用Ortho Vitros 4600检测的分析物为钾(K)、天冬氨酸转氨酶(AST)、乳酸脱氢酶(LDH)、钠(NA)和丙氨酸转氨酶(ALT)。样本分为两组:一组垂直放置在架子上,另一组水平放置在实验室快递员使用的信使包式软边冷藏箱中。在袋子中放置一个冷敷袋,与两组样本等距。为模拟快递运输,将袋子携带30分钟,返回实验室并立即进行分析。随后,对样本重新离心并再次分析。使用GraphPad Prism进行统计分析。垂直和水平放置的样本在运输后LDH均升高,尽管重新离心后仍保持升高。重新离心后,仅水平放置的样本中K升高。值得注意的是,AST以及WBC、RBC和PLT在运输后均升高,但重新离心后降至基线浓度。在快递运输过程中,离心后的PST应保持垂直放置。运输后,对于某些化学检测,可能需要对血浆凝胶管重新离心。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验