Clin Chim Acta. 2019 Sep;496:18-24. doi: 10.1016/j.cca.2019.06.013. Epub 2019 Jun 12.
The BD Barricor tube uses a novel mechanical separator designed to eliminate gel artifacts, decrease cellular contamination, and improve stability. Here, we evaluated the Barricor tube as a possible replacement for PST using Beckman Coulter analyzers under both optimal, alternative, and suboptimal centrifugation conditions based on BD recommendations.
Paired PST and Barricor samples were collected from 4 local hospitals and processed based on site-specific preanalytical systems involving automated or manual centrifugation. Centrifugation conditions ranged from 1912 ×g for 10 min (suboptimal), 2060 g for 10 min (alternative), and 4000 ×g for 3 or 10 min (optimal). Tube volume (4.5 vs. 5.5 ml) was also assessed. Forty-three chemistry and immunochemistry analytes were measured on Beckman Coulter DxC and DxI analyzers.
Using an automated preanlaytical system with suboptimal spin conditions, no bias between PST and Barricor was observed for all analytes tested except lactate dehydrogenase (LD). Further investigation revealed significant increase in LD when Barricor was spun for 10 min at 1912, 2060 and 4000 ×g, ranging from +7.4-19.4% vs. PST across the entire measurement interval (87-493 U/l). Smaller tube volume was also associated with higher LD. Differences in LD occurred despite no change in other hemolysis markers such as potassium, phosphate, and AST.
LD is most sensitive to varying centrifugation conditions (time and speed) in Barricor tubes. We recommend that BD centrifugation protocols should be closely evaluated to determine if Barricor is equivalent to PST under local preanalytical configurations.
BD Barricor 管采用了一种新颖的机械分离器,旨在消除凝胶伪影、减少细胞污染并提高稳定性。在这里,我们根据 BD 的建议,在最优、替代和次优离心条件下,使用贝克曼库尔特(Beckman Coulter)分析仪评估了 Barricor 管作为 PST 的替代物的可能性。
从当地的 4 家医院采集了配对的 PST 和 Barricor 样本,并根据特定于地点的预分析系统进行处理,该系统涉及自动或手动离心。离心条件范围从 1912×g 离心 10 分钟(次优)、2060g 离心 10 分钟(替代)和 4000×g 离心 3 或 10 分钟(最优)。还评估了管体积(4.5 与 5.5ml)。在贝克曼库尔特 DxC 和 DxI 分析仪上测量了 43 种化学和免疫化学分析物。
使用自动化的预分析系统和次优的旋转条件,除乳酸脱氢酶(LD)外,所有测试的分析物在 PST 和 Barricor 之间没有观察到偏差。进一步的研究表明,当 Barricor 在 1912、2060 和 4000×g 下旋转 10 分钟时,LD 显著增加,与整个测量区间内的 PST 相比,范围为+7.4-19.4%(87-493U/l)。较小的管体积也与较高的 LD 相关。尽管钾、磷酸盐和 AST 等其他溶血标志物没有变化,但 LD 存在差异。
LD 对 Barricor 管中变化的离心条件(时间和速度)最为敏感。我们建议应仔细评估 BD 离心方案,以确定在当地预分析配置下,Barricor 是否与 PST 等效。