Tsanakas J N, Wilson A J, Boon A W
Arch Dis Child. 1987 May;62(5):506-8. doi: 10.1136/adc.62.5.506.
The reproducibility of the output of seven different nebulisers was tested. Nebulisers with a minimal increase in output at higher flow rates had less variability in the output. The selection of a nebuliser for bronchial provocation tests depends on the intranebuliser variability. Two nebulisers had highly reproducible outputs.
对七种不同雾化器的输出再现性进行了测试。在较高流速下输出增加最小的雾化器,其输出的变异性较小。用于支气管激发试验的雾化器的选择取决于雾化器内部的变异性。有两种雾化器具有高度可重复的输出。