Suppr超能文献

在瓣膜置换期间及置换后即刻对Björk-Shiley和Medtronic Hall瓣膜进行瓣环固定。

Disc immobilization of Björk-Shiley and Medtronic Hall valves during and immediately after valve replacement.

作者信息

Pai G P, Ellison R G, Rubin J W, Moore H V, Kamath M V

出版信息

Ann Thorac Surg. 1987 Jul;44(1):73-6. doi: 10.1016/s0003-4975(10)62361-2.

Abstract

During the last fourteen years, 377 unileaflet tilting-disc prosthetic valves (Björk-Shiley and Medtronic Hall) have been used for single or multiple valve replacements with and without concomitant coronary artery by-pass grafting. In the past five years, five instances of disc immobilization (three in the mitral and two in the aortic position) occurred either at the time of weaning from cardiopulmonary bypass or immediately thereafter. When the implanted site of the prosthetic valve was the mitral position, reexploration in 2 patients revealed chordal remnants in the subannular area stuck between the disc occluder and the valve ring, thereby immobilizing the disc. In the third instance, the free movement of the disc was impeded by the left ventricular myocardium. In the aortic position, an unraveled suture impacted between the disc occluder and the valve ring immobilized the disc in 1 patient. In the other patient, the cause of the malfunction could not be determined at the time of reexploration. The 1 death among these 5 patients was directly related to the malfunction of the prosthesis. The mechanism, recognition, treatment, and prevention of this catastrophic malfunction of tilting-disc valves are discussed.

摘要

在过去的十四年里,377个单叶倾斜盘式人工瓣膜(Björk-Shiley和Medtronic Hall型)被用于单瓣膜或多瓣膜置换手术,部分手术还同时进行了冠状动脉搭桥术。在过去五年中,有五例瓣膜盘固定事件发生(三例在二尖瓣位,两例在主动脉瓣位),均发生在体外循环撤机时或之后不久。当人工瓣膜植入部位为二尖瓣位时,对2例患者再次手术发现瓣环下区域的腱索残端卡在瓣膜盘封堵器和瓣环之间,从而使瓣膜盘固定。第三例中,瓣膜盘的自由活动受到左心室心肌的阻碍。在主动脉瓣位,1例患者的瓣膜盘封堵器和瓣环之间夹有一根松开的缝线,导致瓣膜盘固定。另一例患者在再次手术时无法确定故障原因。这5例患者中有1例死亡与人工瓣膜故障直接相关。本文讨论了倾斜盘式瓣膜这种灾难性故障的机制、识别、治疗及预防。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验