Center of Digital Dentistry, Peking University School and Hospital of Stomatology, Beijing, China; Faculty of Prosthodontics, Peking University School and Hospital of Stomatology, Beijing, China; National Engineering Research Center of Oral Biomaterials and Digital Medical Devices & National Center of Stomatology & National Clinical Research Center for Oral Diseases & Beijing Key Laboratory of Digital Stomatology & Research Center of Engineering and Technology for Digital Dentistry of Ministry of Health, Beijing, China.
Center of Digital Dentistry, Peking University School and Hospital of Stomatology, Beijing, China; National Engineering Research Center of Oral Biomaterials and Digital Medical Devices & National Center of Stomatology & National Clinical Research Center for Oral Diseases & Beijing Key Laboratory of Digital Stomatology & Research Center of Engineering and Technology for Digital Dentistry of Ministry of Health, Beijing, China.
J Dent. 2022 Oct;125:104279. doi: 10.1016/j.jdent.2022.104279. Epub 2022 Sep 5.
The purpose of this study was to evaluate a novel 3D-printed custom tray for impressions of edentulous jaws, and to compare it with conventional impression trays.
Fifteen edentulous patients were enroled to evaluate the accuracy and border extension of a novel custom tray in the shape of a complete denture. Four impressions were made for each patient. Impressions made using novel custom trays were considered the experimental group, impressions made using conventional custom trays were considered the control group, and impressions made using final dentures and non-pressure custom trays were considered the reference groups. The experimental and control groups were compared with the reference groups using three-dimensional (3D) comparison analysis, and the impressions were further divided into regions. The root mean square (RMS) value was calculated to analyse the differences in impression morphology. Additionally, the experimental and control groups were compared to analyse border extension at standard locations.
Compared to the final denture impression as a reference, the diagnostic denture impression (RMS:0.146 ± 0.024 mm) was closer to the reference than the conventional impression (RMS:0.176 ± 0.047 mm), with a significant difference only in the secondary stress-bearing area. The border extension of the diagnostic denture impression was slightly longer than the conventional impression; however, the difference was not statistically significant.
The impressions made using the novel custom tray were similar to those made with a definitive complete denture. However, no significant differences were noted when compared with the conventional impressions.
本研究旨在评估一种用于无牙颌印模的新型 3D 打印定制托盘,并将其与传统印模托盘进行比较。
本研究纳入了 15 名无牙颌患者,以评估一种全口义齿形状的新型定制托盘的准确性和边缘伸展性。每位患者进行了 4 次印模。使用新型定制托盘进行的印模被视为实验组,使用传统定制托盘进行的印模被视为对照组,使用最终义齿和非压力定制托盘进行的印模被视为参考组。使用三维(3D)比较分析比较实验组和对照组与参考组之间的差异,并进一步将印模分为区域。计算均方根(RMS)值来分析印模形态的差异。此外,还比较了实验组和对照组在标准位置的边缘伸展情况。
与最终义齿印模作为参考相比,诊断义齿印模(RMS:0.146±0.024mm)更接近参考,而常规印模(RMS:0.176±0.047mm)则有显著差异,仅在次要承重区域。诊断义齿印模的边缘伸展略长于常规印模,但差异无统计学意义。
使用新型定制托盘进行的印模与使用最终全口义齿进行的印模相似。然而,与传统印模相比,没有发现显著差异。