J Refract Surg. 2022 Sep;38(9):565-571. doi: 10.3928/1081597X-20220802-03. Epub 2022 Sep 1.
To compare the accuracy of the Barrett Integrated K (IK) toric calculator with the standard Barrett toric calculator.
Consecutive patients who underwent cataract extraction with implantation of a toric intraocular lens at the Rabin Medical Center, Israel, were reviewed. Errors in predicted postoperative refractive astigmatism were calculated for the Barrett toric calculator using biometry measurements only and with the IK tool using biometry and tomography. Both methods were assessed with predicted and measured posterior corneal astigmatism (PPCA and MPCA, respectively).
The study included 73 eyes of 59 patients. The mean centroid prediction error using PPCA (0.08 ± 0.80 D @ 78°) was significantly different compared with MPCA (0.07 ± 0.80 D @ 48°, = .016). In addition, a significant difference between IK-PPCA (0.06 ± 0.80 D @ 80°) and IK-MPCA (0.05 ± 0.80 D @ 38°) was found ( = .023). The median absolute prediction error ranged from 0.55 D using IK-PPCA to 0.60 D using PPCA, with no significant differences between the four calculation versions. No significant differences were found between the calculators in the predictability rates within ±0.50, ±0.75, and ±1.00 D. Analysis of one eye of each patient showed similar results.
The IK calculator yielded comparable outcomes to the standard Barrett calculator. Although differences in the mean centroid errors were found, they were clinically insignificant and predominantly seen in the axis of the predicted astigmatism error. These minor differences were mainly attributed to the incorporation of the MPCA in the calculation. .
比较 Barrett Integrated K(IK)散光计算器与标准 Barrett 散光计算器的准确性。
回顾在以色列拉宾医学中心接受白内障超声乳化吸除联合散光型人工晶状体植入术的连续患者。仅使用生物测量值和使用 IK 工具(同时使用生物测量和断层扫描)计算 Barrett 散光计算器预测术后屈光性散光的误差。分别使用预测的和测量的后角膜散光(PPCA 和 MPCA)评估这两种方法。
本研究纳入了 59 名患者的 73 只眼。使用 PPCA 时的中心点预测误差的平均值(0.08 ± 0.80 D @ 78°)与 MPCA(0.07 ± 0.80 D @ 48°, =.016)有显著差异。此外,IK-PPCA(0.06 ± 0.80 D @ 80°)和 IK-MPCA(0.05 ± 0.80 D @ 38°)之间也存在显著差异( =.023)。使用 IK-PPCA 时的中位绝对预测误差范围为 0.55 D,使用 PPCA 时为 0.60 D,四种计算版本之间无显著差异。四个计算器在预测精度方面(±0.50、±0.75 和 ±1.00 D 内)没有显著差异。对每位患者的一只眼进行分析,结果相似。
IK 计算器与标准 Barrett 计算器的结果相当。尽管在中心点误差平均值上存在差异,但差异在临床上并不显著,主要见于预测散光误差的轴位。这些小的差异主要归因于 MPCA 纳入到计算中。