PET ECG 门控在 PET/MR 成像评估心脏功能的交叉验证研究中的价值。

Value of PET ECG gating in a cross-validation study of cardiac function assessment by PET/MR imaging.

机构信息

Nuklearmedizinische Klinik und Poliklinik, Klinikum Rechts der Isar, School of Medicine, Technical University of Munich, Munich, Germany.

Klinik und Poliklinik für Innere Medizin I, Klinikum Rechts der Isar, School of Medicine, Technical University of Munich, Munich, Germany.

出版信息

J Nucl Cardiol. 2023 Jun;30(3):1050-1060. doi: 10.1007/s12350-022-03105-2. Epub 2022 Sep 30.

Abstract

BACKGROUND

This work investigated the impact of different cardiac gating methods on the assessment of cardiac function by FDG-PET in a cross-validation PET/MR study.

METHODS AND RESULTS

MR- and PET-based left ventricular end-diastolic, end-systolic volumes, and ejection fraction (EDV, ESV, and EF) were delineated in 30 patients with a PET/MR examination. Cardiac PET imaging was performed using three ECG gating methods: fixed number of gates per beat (STD), STD with a beat acceptance window (STD-BR), and fixed gate duration (FW). High MR-PET correlations were found in all the values. ESVs correlated better than EDVs and EFs: Pearson's r coefficient [0.92, 0.92, 0.92] in ESV vs [0.75, 0.81, 0.80] in EDV and [0.79, 0.91, 0.87] in EF, for each method [STD, STD-BR, FW]. Biases with respect to MRI for all the evaluated PET methods were less than 13% in EDV, 5% in ESV, and 14% in EF, but with wide limits of agreements, in the range (59-68)% in EDV, (65-70)% in ESV, and (49-71)% in EF. STD showed the strongest disagreement, while there were no marked differences between STD-BR and FW.

CONCLUSION

Based on these findings, PET- and MR-based cardiac function parameters were highly correlated but in substantial disagreement with variabilities introduced by the selected PET ECG gating method. The most significant differences were associated with the ECG gating method susceptible to highly irregular beats, while similar performance was observed in the methods using uniform adjustment of gates width per beat with the beat acceptance window, and fixed gate width along all the beats. Thus, strict quality controls of R peak detection are needed to minimize its impact on the function assessment.

摘要

背景

本研究通过 PET/MR 交叉验证研究,探讨了不同心脏门控方法对 FDG-PET 评估心脏功能的影响。

方法和结果

对 30 例患者的 PET/MR 检查进行了 MR 和 PET 左心室舒张末期、收缩末期容积和射血分数(EDV、ESV 和 EF)的勾画。心脏 PET 成像采用三种 ECG 门控方法:每搏固定门数(STD)、STD 加搏门宽接受窗(STD-BR)和固定门宽(FW)。所有值均显示出较高的 MR-PET 相关性。ESV 比 EDV 和 EF 相关性更好:每个方法的 Pearson r 系数分别为 [0.92,0.92,0.92] 与 ESV [0.75,0.81,0.80] 与 EDV 和 [0.79,0.91,0.87] 与 EF。对于所有评估的 PET 方法,相对于 MRI 的偏差在 EDV 中小于 13%,在 ESV 中小于 5%,在 EF 中小于 14%,但一致性范围很宽,在 EDV 中为(59-68)%,在 ESV 中为(65-70)%,EF 中为(49-71)%。STD 显示出最强的不一致性,而 STD-BR 和 FW 之间没有明显差异。

结论

基于这些发现,基于 PET 和 MR 的心脏功能参数高度相关,但与所选 PET ECG 门控方法引入的变异性存在很大差异。差异与易受不规则搏动影响的 ECG 门控方法最相关,而在使用均匀调整每搏门宽与搏门宽接受窗以及固定沿所有搏动的门宽的方法中观察到相似的性能。因此,需要严格控制 R 波检测的质量,以最大程度地减少其对功能评估的影响。