Acta Chim Slov. 2022 Sep 26;69(3):507-518. doi: 10.17344/acsi.2022.7348.
A sensitive and inexpensive differential pulse voltammetric technique was applied to investigate the electrochemical behavior of alfuzosin hydrochloride at two different working electrodes: silica gel modified carbon paste and pencil graphite electrodes (PGE). The voltammetric conditions were optimized using cyclic voltammetry, showing an irreversible anodic peak in Britton-Robinson buffered medium (pH 6) at 0.86-0.90 V. The electrochemical responses were linearly correlated with alfuzosin concentrations (R2> 0.999) in the ranges of 0.6-20 and 0.3-20 μM, exhibiting higher electrocatalytic activity at PGE with a low detection limit/ detectability of 0.099 μM. In addition, this study was a successful attempt for the drug determination in tablets and spiked urine samples with green profile evaluation, employing the National Environmental Methods Index, analytical Eco-Scale score, and Green Analytical Procedure Index.
硅胶修饰碳糊电极和铅笔芯石墨电极(PGE)中的电化学行为。通过循环伏安法优化了伏安条件,在 Britton-Robinson 缓冲介质(pH 6)中于 0.86-0.90 V 处显示出不可逆的阳极峰。电化学响应与盐酸阿夫唑嗪浓度(R2>0.999)在 0.6-20 和 0.3-20 μM 范围内呈线性相关,在 PGE 上表现出更高的电催化活性,检测限/检出限低至 0.099 μM。此外,这项研究在采用国家环境方法索引、分析生态标度评分和绿色分析程序索引的情况下,在片剂和加标尿液样品中成功尝试了药物测定,具有绿色分析的特点。