Metabolic and Newborn Screening Laboratory, Department of Pediatrics, University of Szeged, Korányi fasor 14-15, Szeged H-6720, Hungary.
Bruker BioSpin GmbH, Rudolf-Plank-Straße 23, Ettlingen D-76275, Germany.
Clin Biochem. 2023 Jan;111:72-80. doi: 10.1016/j.clinbiochem.2022.09.010. Epub 2022 Oct 3.
Determination of methylmalonic acid (MMA) from dried blood spots (DBS) is commonly performed in clinical diagnostics and newborn screening for propionic acidemia (PA) and methylmalonic acidemia. Isobaric compounds of MMA having the same mass can affect diagnostic reliability and quantitative results, which represents a previously unrecognized pitfall in clinical assays for MMA. We set out to identify interfering substances of MMA in DBS, serum and urine samples from confirmed patients with PA and methylmalonic acidemia.
Techniques included quadrupole time-of-flight high-resolution mass spectrometry (QTOF HR-MS), nuclear magnetic resonance (NMR) spectroscopy, liquid chromatography (LC) and tandem mass spectrometry (MS/MS).
The five isobaric metabolites detected in DBS, serum and urine from PA and methylmalonic acidemia patients were confirmed as 2-methyl-3-hydroxybutyrate, 3-hydroxyisovalerate, 2-hydroxyisovalerate, 3-hydroxyvalerate and succinate using a series of experiments. An additional unknown substance with low abundance remained unidentified.
The presented results facilitate the diagnostic and quantitative reliability of the MMA determination in clinical assays. Isobaric species should be investigated in assays for MMA to eliminate possible interference in a wide range of conditions including PA, methylmalonic acidemia, a vitamin B deficiency, ketosis and lactic acidosis.
从干血斑 (DBS) 中测定甲基丙二酸 (MMA) 通常用于丙酸血症 (PA) 和甲基丙二酸血症的临床诊断和新生儿筛查。具有相同质量的 MMA 的等摩尔化合物会影响诊断的可靠性和定量结果,这代表了 MMA 临床检测中以前未被认识到的陷阱。我们旨在确定 DBS、血清和尿液样本中与 PA 和甲基丙二酸血症相关的 MMA 干扰物质。
采用四级杆飞行时间高分辨率质谱 (QTOF HR-MS)、核磁共振 (NMR) 光谱、液相色谱 (LC) 和串联质谱 (MS/MS) 等技术。
通过一系列实验,确定从 PA 和甲基丙二酸血症患者的 DBS、血清和尿液中检测到的五种等摩尔代谢物分别为 2-甲基-3-羟基丁酸、3-羟基异戊酸、2-羟基异戊酸、3-羟基戊酸和琥珀酸。一种丰度较低的未知物质仍未被鉴定。
本研究结果提高了 MMA 测定在临床检测中的诊断和定量可靠性。在 MMA 检测中应研究等摩尔物质,以消除包括 PA、甲基丙二酸血症、维生素 B 缺乏症、酮症和乳酸酸中毒在内的广泛条件下可能存在的干扰。