Gnerucci A, Esposito M, Ghirelli A, Pini S, Paoletti L, Barca R, Fondelli S, Alpi P, Grilli B, Rossi F, Scoccianti S, Russo S
Department of Physics and Astronomy, University of Florence, Florence, Italy.
Medical Physics Unit, Azienda USL Toscana Centro, Florence, Italy.
Strahlenther Onkol. 2023 Jan;199(1):55-66. doi: 10.1007/s00066-022-02008-y. Epub 2022 Oct 13.
To compare two left breast cancer patient cohorts (tangential vs. locoregional deep-inspiration breath-hold - DIBH treatment) with different predefined beam gating thresholds and to evaluate their impact on motion management and DIBH stability.
An SGRT-based clinical workflow was adopted for the DIBH treatment. Intrafractional monitoring was performed by tracking both the respiratory signal and the real-time displacement between the isocenter on the daily reference surface and on the live surface ("SGRT shift"). Beam gating tolerances were 5 mm/4 mm for the SGRT shifts and 5 mm/3 mm for the gating window amplitude for breast tangential and breast + lymph nodes locoregional treatments, respectively. A total of 24 patients, 12 treated with a tangential technique and 12 with a locoregional technique, were evaluated for a total number of 684 fractions. Statistical distributions of SGRT shift and respiratory signal for each treatment fraction, for each patient treatment, and for the two population samples were generated.
Lateral cumulative distributions of SGRT shifts for both locoregional and tangential samples were consistent with a null shift, whereas longitudinal and vertical ones were slightly negative (mean values < 1 mm). The distribution of the percentage of beam on time with SGRT shift > 3 mm, > 4 mm, or > 5 mm was extended toward higher values for the tangential sample than for the locoregional sample. The variability in the DIBH respiration signal was significantly greater for the tangential sample.
Different beam gating thresholds for surface-guided DIBH treatment of left breast cancer can impact motion management and DIBH stability by reducing the frequency of the maximum SGRT shift and increasing respiration signal stability when tighter thresholds are adopted.
比较两个左乳腺癌患者队列(切线野与局部区域深吸气屏气 - DIBH治疗),其采用不同的预定义射束门控阈值,并评估它们对运动管理和DIBH稳定性的影响。
DIBH治疗采用基于表面引导放疗(SGRT)的临床工作流程。通过跟踪呼吸信号以及每日参考表面与实时表面上的等中心之间的实时位移(“SGRT位移”)进行分次内监测。对于乳腺切线野治疗和乳腺 + 淋巴结局部区域治疗,SGRT位移的射束门控公差分别为5毫米/4毫米和5毫米/3毫米,门控窗口幅度的公差分别为5毫米/3毫米。总共评估了24例患者,其中12例采用切线技术治疗,12例采用局部区域技术治疗,共计684个分次。生成了每个治疗分次、每位患者治疗以及两个人群样本的SGRT位移和呼吸信号的统计分布。
局部区域和切线野样本的SGRT位移的横向累积分布与零位移一致,而纵向和垂直方向的累积分布略为负向(平均值 < 1毫米)。对于切线野样本,SGRT位移 > 3毫米、> 4毫米或 > 5毫米时射束开启时间百分比的分布向更高值扩展,高于局部区域样本。切线野样本的DIBH呼吸信号变异性显著更大。
对于左乳腺癌的表面引导DIBH治疗,不同的射束门控阈值可通过降低最大SGRT位移的频率并在采用更严格的阈值时提高呼吸信号稳定性,从而影响运动管理和DIBH稳定性。