Department of Radiation Oncology, Comprehensive Cancer Center, Medical University of Vienna/AKH Wien, Währinger Gürtel 18-20, Vienna, 1090, Austria.
Strahlenther Onkol. 2024 Apr;200(4):306-313. doi: 10.1007/s00066-023-02153-y. Epub 2023 Oct 5.
To investigate the feasibility of a thermo-optical surface imaging (SGRT) system combined with room-based stereoscopic X‑ray image guidance (IGRT) in a dedicated breast deep inspiration breath-hold (DIBH) irradiation workflow. In this context, benchmarking of portal imaging (EPID) and cone-beam CT (CBCT) against stereoscopic X‑rays was performed.
SGRT + IGRT data of 30 left-sided DIBH breast patients (1 patient with bilateral cancer) treated in 351 fractions using thermo-optical surface imaging and X-ray IGRT were retrospectively analysed. Patients were prepositioned based on a free-breathing surface reference derived from a CT scan. Once the DIBH was reached using visual feedback, two stereoscopic X‑ray images were acquired and registered to the digitally reconstructed radiographs derived from the DIBH CT. Based on this registration, a couch correction was performed. Positioning and monitoring by surface and X-ray imaging were verified by protocol-based EPID or CBCT imaging at selected fractions and the calculation of residual geometric deviations.
The median X‑ray-derived couch correction vector was 4.9 (interquartile range [IQR] 3.3-7.1) mm long. Verification imaging was performed for 134 fractions (216 RT field verifications) with EPID and for 37 fractions with CBCT, respectively. The median 2D/3D deviation vector length over all verification images was 2.5 (IQR 1.6-3.9) mm/3.4 (IQR 2.2-4.8) mm for EPID/CBCT, both being well within the planning target volume (PTV) margins (7 mm). A moderate correlation (0.49-0.65) was observed between the surface signal and X-ray position in DIBH.
DIBH treatments using thermo-optical SGRT and X-ray IGRT were feasible for breast cancer patients. Stereoscopic X‑ray positioning was successfully verified by standard IGRT techniques.
研究热光表面成像(SGRT)系统与基于房间的立体 X 射线图像引导(IGRT)在专用乳房深吸气屏气(DIBH)照射工作流程中的可行性。在此背景下,对门成像(EPID)和锥形束 CT(CBCT)与立体 X 射线进行了基准测试。
回顾性分析了 30 例左侧 DIBH 乳腺癌患者(1 例双侧癌症患者)的 351 个分次的 SGRT+IGRT 数据,这些患者使用热光表面成像和 X 射线 IGRT 进行治疗。患者根据 CT 扫描得出的自由呼吸表面参考进行预定位。一旦通过视觉反馈达到 DIBH,就会采集两个立体 X 射线图像,并将其与来自 DIBH CT 的数字重建射线照相进行注册。基于此注册,对治疗床进行校正。通过基于协议的 EPID 或 CBCT 成像在选定的分次中验证和监测表面和 X 射线成像的定位和监测,并计算残余几何偏差。
X 射线衍生的治疗床校正向量的中位数为 4.9(四分位距[IQR] 3.3-7.1)mm。分别对 134 个分次(216 个 RT 场验证)进行了 EPID 验证成像,对 37 个分次进行了 CBCT 验证成像。所有验证图像的中位 2D/3D 偏差向量长度分别为 2.5(IQR 1.6-3.9)mm/3.4(IQR 2.2-4.8)mm,均在计划靶区(PTV)边界(7mm)内。在 DIBH 中,观察到表面信号与 X 射线位置之间存在中度相关性(0.49-0.65)。
使用热光 SGRT 和 X 射线 IGRT 的 DIBH 治疗对乳腺癌患者是可行的。立体 X 射线定位可通过标准 IGRT 技术成功验证。