MacCrehan W A, Schönberger E
J Chromatogr. 1987 Jun 5;417(1):65-78. doi: 10.1016/0378-4347(87)80092-0.
Baseline separation of the isomers of retinol using reversed-phase high-performance liquid chromatography (HPLC) in less than 30 min is presented. A new approach to the detection of retinol using electrochemical detection is developed. The oxidative electrochemistry of retinol is studied at a glassy-carbon electrode using coulometry, ultraviolet-visible spectrophotometry and HPLC. Amperometric detection in HPLC for retinol provided a linear response from 0 to 1.5 micrograms/ml and a detection limit of 4.1 ng/ml. Electrochemical detection was compared to ultraviolet-visible absorbance detection for the determination of retinol in human serum extracts. Good agreement is found for the results obtained with the two detectors.
本文介绍了使用反相高效液相色谱法(HPLC)在不到30分钟的时间内实现视黄醇异构体的基线分离。开发了一种使用电化学检测法检测视黄醇的新方法。利用库仑法、紫外-可见分光光度法和HPLC在玻碳电极上对视黄醇的氧化电化学进行了研究。HPLC中视黄醇的安培检测在0至1.5微克/毫升范围内提供线性响应,检测限为4.1纳克/毫升。将电化学检测与紫外-可见吸光度检测用于测定人血清提取物中的视黄醇进行了比较。两种检测器获得的结果具有良好的一致性。