Computed tomography based assessment of programmable shunt valve settings.

作者信息

Decramer Thomas, Smeijers Steven, Vanhoyland Michaël, Coudyzer Walter, Van Calenbergh Frank, van Loon Johannes, De Vloo Philippe, Theys Tom

机构信息

Department of Neurosurgery, University Hospitals Leuven, Leuven, Belgium.

Research Group Experimental Neurosurgery and Neuroanatomy, KU Leuven and the Leuven Brain Institute, Leuven, Belgium.

出版信息

Brain Spine. 2021 Aug 6;1:100003. doi: 10.1016/j.bas.2021.100003. eCollection 2021.

Abstract

INTRODUCTION

Programmable shunt valve settings can sometimes be difficult to assess using classic read-out tools, warranting a skull X-ray.

RESEARCH QUESTION

Can we use available head computed tomography (CT) scans to determine the valve settings, in order to obviate the need for additional skull X-rays?

MATERIAL AND METHODS

The valve setting of two different programmable shunts (Codman Certas Plus® and Sophysa Polaris®) were assessed by two blinded observers in 24 patients using 65 head CT scans (slice thickness ≤2 ​mm). Using multi-planar reconstruction (MPR) tools, images were resliced according to the direction of the valve, allowing a direct readout of the valve settings. We validated our CT based method against 32 available skull X-rays.

RESULTS

For all CT scans it was possible to assess the valve setting. No interobserver variability was found and there was a 100 ​% concordance between the CT based method and skull X-rays.

DISCUSSION

CT based assessment of programmable shunt valve settings is feasible and reliable. It may obviate the need for additional skull x-rays when a head CT scan is available.

CONCLUSIONS

This technique can reduce radiation exposure and can be applied to historical CT imaging with unknown valve settings.

摘要

引言

使用传统读出工具有时难以评估可编程分流阀的设置,因此需要进行颅骨X线检查。

研究问题

我们能否使用现有的头部计算机断层扫描(CT)来确定阀门设置,从而无需额外的颅骨X线检查?

材料与方法

两名盲法观察者使用65例头部CT扫描(层厚≤2毫米)对24例患者的两种不同可编程分流器(Codman Certas Plus®和Sophysa Polaris®)的阀门设置进行评估。使用多平面重建(MPR)工具,根据阀门方向重新切片图像,从而直接读出阀门设置。我们将基于CT的方法与32例现有的颅骨X线检查结果进行了验证。

结果

对于所有CT扫描,均能够评估阀门设置。未发现观察者间差异,基于CT的方法与颅骨X线检查结果的一致性为100%。

讨论

基于CT评估可编程分流阀设置是可行且可靠的。当有头部CT扫描时,可能无需额外的颅骨X线检查。

结论

该技术可减少辐射暴露,并可应用于阀门设置未知的既往CT成像。