Ono Tomohiro, Kawata Kohei, Nakamura Mitsuhiro, Uto Megumi, Mizowaki Takashi
Department of Radiation Oncology and Image-Applied Therapy, Graduate School of Medicine, Kyoto University, 54 Kawahara-Cho, Shogoin, Sakyo-Ku, Kyoto, Kyoto, 606-8507, Japan.
Division of Clinical Radiology Service, Kyoto University Hospital, 54 Kawahara-Cho, Shogoin, Sakyo-Ku, Kyoto, Kyoto, 606-8507, Japan.
Radiol Phys Technol. 2023 Mar;16(1):10-19. doi: 10.1007/s12194-022-00684-0. Epub 2022 Oct 22.
Recently, the International Atomic Energy Agency and the American Association of Physicists in Medicine reported correction factors (CFs) for detector-response variation considering the uncertainty in detector readings in small-field dosimetry. In this study, the effect of CFs on small-field dosimetry measurements was evaluated for single-isocenter stereotactic radiotherapy for brain metastases. The output factors (OPFs) were measured with and without CFs in a water-equivalent sphere phantom using TrueBeam with a flattening-filter-free energy of 10 MV. Five detectors were used in a perpendicular orientation: CC01, 3D pinpoint ionization chambers, unshielded SFD detector, shielded EDGE detector, and microDiamond detector. First, the square-field sizes were set to 5-100 mm using a multi-leaf collimator (MLC) field. The OPFs were evaluated in the presence and absence of CFs. Second, single-isocenter stereotactic irradiation was performed on 22 brain metastases in 15 patients following dynamic conformal arc (DCA) treatment. The equivalent field size was calculated using the MLC aperture for each planning target volume. For the OPFs, the mean deviations from the median of the doses measured with detectors other than the CC01 for square-field sizes larger than 10 mm were within ± 4.3% of the median without CFs, and ± 3.3% with CFs. For DCA plans, the deviations without and with CFs were - 2.3 ± 1.9% and - 4.8 ± 2.4% for CC01, - 1.1 ± 3.0% and 1.0 ± 1.6% for 3D pinpoint, 8.8 ± 3.0% and 2.9 ± 2.8% for SFD, - 3.1 ± 3.0% and - 13.5 ± 4.0% for EDGE, and 8.9 ± 2.1% and 0.8 ± 1.9% for microDiamond. This feasibility study confirmed that the deviation of the detectors can be reduced using an appropriate detector with CFs.
最近,国际原子能机构和美国医学物理师协会报告了考虑小射野剂量测定中探测器读数不确定性的探测器响应变化校正因子(CFs)。在本研究中,评估了CFs对脑转移瘤单等中心立体定向放射治疗中小射野剂量测定测量的影响。在具有10 MV无均整器能量的TrueBeam直线加速器上,在水等效球模体中使用和不使用CFs测量输出因子(OPFs)。使用五个垂直方向的探测器:CC01、3D针点电离室、无屏蔽SFD探测器、屏蔽EDGE探测器和微金刚石探测器。首先,使用多叶准直器(MLC)射野将方野尺寸设置为5 - 100 mm。在有和没有CFs的情况下评估OPFs。其次,对15例患者中的22个脑转移瘤进行动态适形弧(DCA)治疗后进行单等中心立体定向照射。使用每个计划靶体积的MLC孔径计算等效野尺寸。对于OPFs,对于大于10 mm的方野尺寸,使用除CC01之外的探测器测量的剂量中位数的平均偏差在不使用CFs时为中位数的±4.3%以内,使用CFs时为±3.3%以内。对于DCA计划,对于CC01,不使用和使用CFs时的偏差分别为 - 2.3±1.9%和 - 4.8±2.4%,对于3D针点电离室分别为 - 1.1±3.0%和1.0±1.6%,对于SFD分别为8.8±3.0%和2.9±2.8%,对于EDGE分别为 - 3.1±3.0%和 - 13.5±4.0%,对于微金刚石分别为8.9±2.1%和0.8±1.9%。这项可行性研究证实,使用带有CFs的合适探测器可以减少探测器的偏差。