Gullberg G T, Wehrli F W, Shimakawa A, Simons M A
Radiology. 1987 Oct;165(1):241-6. doi: 10.1148/radiology.165.1.3628776.
The authors present a method for obtaining magnetic resonance (MR) images of intra- and extracranial vessels from thin contiguous transaxial sections. A section-selective gradient refocusing pulse sequence with a short repetition time caused flow-related enhancement from spins that flowed perpendicular to the transaxial sections. The signal was further enhanced by means of flow compensation gradients to rephase any phase shifts resulting from moving spins in the presence of the imaging gradients. Coronal and sagittal sections, reformatted from multiple transaxial sections, are shown to have excellent vessel contrast without the use of contrast material. These images were obtained in 12 minutes of acquisition time from as many as 60 sections of 3-mm thickness. Such a technique shows significant promise for MR angiography.
作者介绍了一种从连续薄的横断面获取颅内和颅外血管磁共振(MR)图像的方法。具有短重复时间的层面选择梯度重聚焦脉冲序列可使垂直于横断面流动的自旋产生与血流相关的增强信号。通过流动补偿梯度进一步增强信号,以重新相位化在成像梯度存在下移动自旋产生的任何相移。从多个横断面重新格式化得到的冠状面和矢状面图像显示,无需使用造影剂就具有出色的血管对比度。这些图像在12分钟的采集时间内从多达60个3毫米厚的层面获得。这种技术在磁共振血管造影方面显示出巨大的前景。