Yanagisawa Yohei, Kotaki Tomomi, Nakajima Shun, Uesugi Masafumi, Yamazaki Masashi
Department of Emergency and Critical Care Medicine, University of Tsukuba Hospital, 2-1-1 Amakubo, Tsukuba, Ibaraki 305-8576, Japan.
Department of Orthopaedic Surgery, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8575, Japan.
Trauma Case Rep. 2022 Oct 19;42:100725. doi: 10.1016/j.tcr.2022.100725. eCollection 2022 Dec.
Spinopelvic fixation (SPF) is an effective treatment method for vertically unstable pelvic ring fractures with spinopelvic dissociations (Patel et al., 2022). A heavy container fell on a 35-year-old man who was trapped and sustained injuries. His pelvic ring fracture dislocation was identified as AO Classification 61-C2.3 with rotational and vertical unstable pelvic ring; thus, crab-shaped fixation (SPF modification) was performed (Okuda et al., 2019). The pelvic fracture was fused, and the clinical outcome was good with modified Majeed score of 96. However, set-screw loosening was observed during the postoperative course. Reports of implant failures in SPF for unstable pelvic ring fractures commonly occur. However, only a few reports have demonstrated implant failure of crab-shaped fixation. Written informed consent was obtained from the patient for publication of this case report and accompanying images.
脊柱骨盆固定术(SPF)是治疗伴有脊柱骨盆分离的垂直不稳定骨盆环骨折的一种有效方法(帕特尔等人,2022年)。一个重型容器砸在一名35岁男子身上,他被困并受伤。其骨盆环骨折脱位被确定为AO分类61-C2.3型,伴有旋转和垂直不稳定的骨盆环;因此,实施了蟹形固定术(SPF改良术)(奥田等人,2019年)。骨盆骨折得到融合,临床结果良好,改良马吉德评分为96分。然而,在术后过程中观察到固定螺钉松动。关于不稳定骨盆环骨折的SPF植入失败的报道很常见。然而,仅有少数报道显示蟹形固定术的植入失败。已获得患者的书面知情同意,以发表本病例报告及附带图像。