In TAVR, area sizing is used for balloon-expandable (BE) valves, whereas self-expanding valves are sized to annulus perimeter. For BE valves, this seems illogical: these frames force a circular shape even on an ellipsoid annulus. This can potentially lead to relative undersizing when area sizing is being applied. We developed a perimeter-based sizing algorithm to evaluate the safety and feasibility of perimeter sizing for the Myval BE valve.
In this prospective single-center study, 60 patients with severe aortic stenosis treated with the Myval BE valve were included. Perimeter sizing was used with limited oversizing of 3.7% ± 1.3% compared to the annulus perimeter. After TAVR, clinical outcomes were evaluated at 30 days and 1 year. An echocardiographic follow-up took place at 30 days.
At 30 days, the need for PPI and stroke occurred in 2% and 3% of the patients, respectively. Moreover, cardiac death and moderate-severe PVL were absent. At 1-year, cardiac death and stroke were observed in 3% and 8% of the patients, respectively. In 33.3% of the patients, a larger valve size was implanted compared to the valve size calculated by area sizing.
Perimeter sizing with the Myval BE valve leads to substantial use of larger valve sizes and favorable clinical outcomes, with low PPI and the absence of significant PVL. A randomized controlled trial is being planned to prove the superiority of this alternative sizing method.
在 TAVR 中,球囊扩张型(BE)瓣膜采用面积测量法进行瓣环尺寸测量,而自膨胀瓣膜则根据瓣环周长进行测量。对于 BE 瓣膜来说,这种方法似乎不合理:这些框架即使在椭圆形瓣环上也会强制形成圆形。这可能导致在应用面积测量法时出现相对的尺寸过小问题。我们开发了一种基于周长的尺寸测量算法,以评估 Myval BE 瓣膜周长测量的安全性和可行性。
在这项前瞻性单中心研究中,共纳入了 60 例接受 Myval BE 瓣膜治疗的严重主动脉瓣狭窄患者。周长测量法用于瓣膜尺寸测量,与瓣环周长相比,瓣膜的过度扩张限制在 3.7%±1.3%。TAVR 后,在 30 天和 1 年时评估临床结局。在 30 天时进行超声心动图随访。
在 30 天时,分别有 2%和 3%的患者需要使用质子泵抑制剂(PPI)和发生中风。此外,无心脏死亡和中度至重度肺动脉瓣反流(PVL)发生。在 1 年时,分别有 3%和 8%的患者发生心脏死亡和中风。与面积测量法计算的瓣膜尺寸相比,33.3%的患者植入了更大尺寸的瓣膜。
使用 Myval BE 瓣膜进行周长测量会导致大量使用更大尺寸的瓣膜,且临床结局良好,PPI 发生率低,无明显的 PVL。目前正在计划一项随机对照试验,以证明这种替代尺寸测量方法的优越性。