Suppr超能文献

一种应用低温等离子体的新型手术系统的烟雾分析

Smoke analysis of a new surgical system that applies low-temperature plasma.

作者信息

Zhang Boya, Guan Qingyu, Zhu Yunsheng, Zhu Jingjin, Liu Xiaohan, Li Shuaiqi, Yang Rungong, Li Xiru

机构信息

School of Medicine, Nankai University, Tianjin, China.

Department of General Surgery, Chinese People's Liberation Army General Hospital, Beijing, China.

出版信息

Ann Transl Med. 2022 Oct;10(19):1053. doi: 10.21037/atm-22-608.

Abstract

BACKGROUND

The high-frequency electrotome (ES), which is widely used in surgical procedures, generates surgical smoke that is potentially hazardous to operating personnel. Previous research shows that the PlasmaBlade (PB) may be able to overcome this problem. The present study set out to analyze potentially hazardous surgical smoke generated during electrosurgery by the ES, the PB, and. a new surgical system that applies low-temperature plasma, the NTS-100.

METHODS

and healthy porcine models were used to compare volatile organic compounds (VOCs) and particulate matter (PM) in smoke generated by the NTS-100, the PB, and the conventional ES when cutting liver, muscle, and skin and subcutaneous tissues. The detected indexes included the VOCs in surgical smoke, the concentration and percentage of each part, the PM2.5 concentration, the mass of particles, and the diameter distribution of particles.

RESULTS

The smoke generated by the NTS-100 contained fewer hazardous components than that generated by the ES (P<0.05) and a comparable amount to that generated by the PB (P>0.05). The PM2.5 concentration and mass of particles in the smoke generated by the NTS-100 were lower than those with the ES (P<0.05 and P<0.01, respectively) and similar to those with the PB (P>0.05). The NTS-100 generated larger particles than did the ES and the PB (P<0.05).

CONCLUSIONS

Surgical smoke contains harmful VOCs and PM, but the NTS-100 generated less hazardous surgical smoke than did the conventional ES and performed comparably to the PB. Therefore, using the NTS-100 may reduce the potential hazard of surgical smoke to operating room personnel.

摘要

背景

高频电刀(ES)在外科手术中广泛使用,会产生对手术人员有潜在危害的手术烟雾。先前的研究表明,等离子刀(PB)或许能够克服这一问题。本研究旨在分析ES、PB以及一种应用低温等离子体的新型手术系统NTS-100在电外科手术过程中产生的潜在有害手术烟雾。

方法

使用健康猪模型比较NTS-100、PB和传统ES在切割肝脏、肌肉、皮肤及皮下组织时产生的烟雾中的挥发性有机化合物(VOCs)和颗粒物(PM)。检测指标包括手术烟雾中的VOCs、各部分的浓度和百分比、PM2.5浓度、颗粒质量以及颗粒直径分布。

结果

NTS-100产生的烟雾中有害成分比ES产生的少(P<0.05),与PB产生的烟雾量相当(P>0.05)。NTS-100产生的烟雾中PM2.5浓度和颗粒质量低于ES产生的烟雾(分别为P<0.05和P<0.01),与PB产生的烟雾相似(P>0.05)。NTS-100产生的颗粒比ES和PB产生的颗粒大(P<0.05)。

结论

手术烟雾含有有害的VOCs和PM,但NTS-100产生的有害手术烟雾比传统ES少,与PB表现相当。因此,使用NTS-100可能会降低手术烟雾对手术室人员的潜在危害。