Suppr超能文献

利用光纤耦合闪烁体探测器系统进行笔形束扫描质子 FLASH 治疗的时间分辨剂量率测量。

Time-resolved dose rate measurements in pencil beam scanning proton FLASH therapy with a fiber-coupled scintillator detector system.

机构信息

Danish Centre for Particle Therapy, Aarhus University Hospital, Aarhus, Denmark.

Department of Clinical Medicine, Faculty of Health, Aarhus University, Aarhus, Denmark.

出版信息

Med Phys. 2023 Apr;50(4):2450-2462. doi: 10.1002/mp.16156. Epub 2022 Dec 29.

Abstract

BACKGROUND

The spatial and temporal dose rate distribution of pencil beam scanning (PBS) proton therapy is important in ultra-high dose rate (UHDR) or FLASH irradiations. Validation of the temporal structure of the dose rate is crucial for quality assurance and may be performed using detectors with high temporal resolution and large dynamic range.

PURPOSE

To provide time-resolved in vivo dose rate measurements using a scintillator-based detector during proton PBS pre-clinical mouse experiments with dose rates ranging from conventional to UHDR.

METHODS

All irradiations were performed at the entrance plateau of a 250 MeV PBS proton beam. A detector system with four fiber-coupled ZnSe:O inorganic scintillators and 20 μs temporal resolution was used for dose rate measurements. The system was first characterized in terms of precision and stem signal. The detector precision was determined through repeated irradiations with the same field. The stem signal contribution was quantified by irradiating two of the detector probes alongside a bare fiber (fiber without a coupled scintillator). Next, the detector system was calibrated against an ionization chamber (IC) with all four detector probes and the IC placed in a water bath at 2 cm depth. A scan pattern covering 9.6 × 9.6 cm was used. Multiple irradiations with different requested nozzle currents provided instantaneous dose rates at the detector positions in the range of 7-1270 Gy/s. The correspondence of the detector signal (in Volts) to the instantaneous dose rate (in Gy/s) was found. The instantaneous dose rate was calculated from the beam current and the spot-to-detector distance assuming a Gaussian beam profile at distances up to 8 mm from the spot. Afterwards, the calibrated system was used in vivo, in mouse experiments, where mouse legs were irradiated with a constant dose and varying field dose rates of 0.7-87.5 Gy/s. The instantaneous dose rate was measured for each delivered spot and the delivered dose was determined as the integrated instantaneous dose rate. The spot dose profile and PBS dose rate map were calculated. The dose contamination to neighbouring mice were measured together with the upper limit of the dose to the mouse body.

RESULTS

The detectors showed high precision with ≤0.4% fluctuations in the measured dose. The stem signal exceeded 10% for spots <5 mm from the optical fiber and >18 mm from the scintillator. It contributed up to 0.2% to the total dose, which was considered negligible. All four detectors showed a non-linear relation between signal and instantaneous dose rate, which was modelled with a polynomial response function. In the mouse experiments, the measured scintillator dose showed 1.8% fluctuations, independent of the field dose rate. The in vivo measured spot dose profile had tails that deviated from a Gaussian profile with measurable dose contributions from spots up to 85 mm from the detector. Neighbour mouse irradiation contributed ∼1% of the total mouse dose. The upper limit of the mouse body dose was 6% of the mouse leg dose.

CONCLUSIONS

A fiber-coupled inorganic scintillator-based detector system can provide high precision in vivo measurements of the instantaneous dose rate if correction for the non-linear dose rate dependency is applied.

摘要

背景

在超高剂量率 (UHDR) 或 FLASH 照射中,铅笔束扫描 (PBS) 质子治疗的空间和时间剂量率分布很重要。为了保证质量,需要对剂量率的时间结构进行验证,这可以使用具有高时间分辨率和大动态范围的探测器来实现。

目的

在常规剂量率到 UHDR 的范围内,使用基于闪烁体的探测器在进行质子 PBS 临床前小鼠实验时,提供时间分辨的体内剂量率测量。

方法

所有照射均在 250 MeV PBS 质子束的入口平台进行。使用具有四个光纤耦合 ZnSe:O 无机闪烁体和 20 μs 时间分辨率的探测器系统进行剂量率测量。该系统首先在精度和本底信号方面进行了表征。通过对同一照射野进行重复照射来确定探测器的精度。通过对两个探测器探头与裸光纤(没有耦合闪烁体的光纤)一起照射来量化本底信号的贡献。接下来,使用四个探测器探头和置于 2cm 水深的电离室 (IC) 对探测器系统进行校准。使用覆盖 9.6×9.6cm 的扫描模式。通过使用不同的请求喷嘴电流进行多次照射,在探测器位置处提供了 7-1270Gy/s 的瞬时剂量率。发现探测器信号(伏特)与瞬时剂量率(Gy/s)之间存在对应关系。瞬时剂量率是根据束流和光斑到探测器的距离计算的,假设在距离光斑 8mm 以内的高斯束分布。然后,在小鼠实验中,使用校准后的系统对小鼠腿部进行照射,以 0.7-87.5Gy/s 的恒定剂量和变化的野剂量率进行照射。测量了每个照射野的瞬时剂量率,并将所给剂量确定为瞬时剂量率的积分。计算了点剂量分布和 PBS 剂量率图。测量了相邻小鼠的剂量污染以及小鼠身体的最大剂量。

结果

探测器具有很高的精度,测量的剂量波动小于 0.4%。对于距离光纤小于 5mm 或距离闪烁体大于 18mm 的光斑,本底信号超过 10%。它对总剂量的贡献高达 0.2%,可以忽略不计。所有四个探测器都显示出信号和瞬时剂量率之间的非线性关系,这可以通过多项式响应函数来建模。在小鼠实验中,无论场剂量率如何,测量的闪烁体剂量都有 1.8%的波动。体内测量的照射野剂量分布具有尾部,与高斯分布有偏差,从探测器开始的 85mm 范围内都可以测量到剂量。相邻小鼠的照射贡献了总小鼠剂量的约 1%。小鼠身体的最大剂量是小鼠腿部剂量的 6%。

结论

如果对非线性剂量率依赖性进行校正,基于光纤耦合无机闪烁体的探测器系统可以提供高精度的体内瞬时剂量率测量。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验