Houdi A A, Crooks P A, Van Loon G R, Schubert C A
College of Pharmacy, Division of Medicinal Chemistry and Pharmacognosy, University of Kentucky, Lexington 40536-0053.
J Pharm Sci. 1987 May;76(5):398-401. doi: 10.1002/jps.2600760512.
The determination of picomolar levels of histamine and its major metabolite, N tau-methylhistamine, in biological fluids was achieved using reversed-phase liquid chromatography coupled with electrochemical detection. A simple sample purification procedure for blood and urine samples was carried out prior to analysis using an Amberlite CG-50 cation-exchange resin, which afforded an excellent recovery of both compounds.