Comparison of the mechanical properties and surface topography of as-received and as-retrieved Copper-Nickel-Titanium arch wires: An ex-vivo study.

机构信息

Orthodontic Research Center, Shiraz University of Medical Sciences, School of Dentistry, Shiraz, Iran.

Orthodontic Research Center, Shiraz University of Medical Sciences, School of Dentistry, Shiraz, Iran.

出版信息

Int Orthod. 2023 Mar;21(1):100720. doi: 10.1016/j.ortho.2022.100720. Epub 2022 Dec 22.

Abstract

AIM

The mechanical and surface characteristics of Copper NiTi wires following intraoral aging in an important topic. This ex-vivo study is designed to assess the three most important mechanical properties; torsional stiffness, 3-point bending and surface hardness of the as-received Copper NiTi orthodontic archwires and compare them to the retrieved ones.

MATERIAL AND METHODS

In this study, 50 round CuNiTi (0.016 inch) and 50 rectangular (0.018×0.025 inch) CuNiTi orthodontic wires (American Orthodontics, Tanzo Copper Nickel Titanium, Sheboygan, WI) were tested. The groups were as follow: 25 as-received 0.016 inch round CuNiTi archwires (group 1), 25 as-retrieved 0.016 inch round CuNiTi archwires (group 2), 25 as-received 0.018×0.025 inch rectangular CuNiTi archwires (group 3), and 25 as-retrieved 0.018×0.025 inch rectangular CuNiTi archwires (group 4). The physical properties of all of the samples were examined using Vickers hardness, bending and torsional tests. The Kolmogorov-Smirnov test was applied to assess the normal distribution of data. An independent t-test was used to compare the variables between the groups.

RESULTS

The results showed that the hardness does not significantly differ between the as-received and as-retrieved archwires in both 0.016 and 0.018×0.025 inch archwires (P< 0.05). The bending test of the round wires showed that the values of the activation forces in the as-received ones were significantly higher than those of the as-retrieved (P=0.00). The torsional and bending test results of rectangular wires showed that the amounts of forces exerted from the as-retrieved archwires were significantly higher than the new archwires (P=0.00).

CONCLUSIONS

Round CuNiTi archwires could be retied in the succeeding visits. However, the bending activation that declines significantly should be took into consideration. On the other hand, since bending and torsion activation of CuNiTi rectangular archwires increased after intraoral use, these wires should be retied cautiously.

摘要

目的

研究口腔内老化对铜镍钛丝力学和表面特性的影响是一个重要的课题。本体外研究旨在评估三种最重要的机械性能;扭转刚度、三点弯曲和表面硬度,比较收到的铜镍钛正畸弓丝和检索到的弓丝。

材料和方法

本研究中,测试了 50 根圆形 CuNiTi(0.016 英寸)和 50 根矩形(0.018×0.025 英寸)CuNiTi 正畸弓丝(美国 Orthodontics,Tanzo 铜镍钛,Sheboygan,WI)。分组如下:25 根新的 0.016 英寸圆形 CuNiTi 弓丝(组 1)、25 根旧的 0.016 英寸圆形 CuNiTi 弓丝(组 2)、25 根新的 0.018×0.025 英寸矩形 CuNiTi 弓丝(组 3)和 25 根旧的 0.018×0.025 英寸矩形 CuNiTi 弓丝(组 4)。使用维氏硬度、弯曲和扭转试验检查所有样品的物理性能。应用 Kolmogorov-Smirnov 检验评估数据的正态分布。独立 t 检验用于比较组间变量。

结果

结果表明,在 0.016 和 0.018×0.025 英寸弓丝中,新收到和旧收到的弓丝的硬度没有显著差异(P<0.05)。圆形弓丝的弯曲试验表明,新收到的弓丝的激活力值明显高于旧收到的弓丝(P=0.00)。矩形弓丝的扭转和弯曲试验结果表明,旧收到的弓丝施加的力明显高于新弓丝(P=0.00)。

结论

圆形 CuNiTi 弓丝可以在随后的就诊中重新结扎。然而,应该考虑到明显下降的弯曲激活。另一方面,由于 CuNiTi 矩形弓丝在口腔内使用后弯曲和扭转激活增加,这些弓丝应谨慎重新结扎。