Lachenmayr B
Augenklinik, Universität München.
Klin Monbl Augenheilkd. 1987 Jul;191(1):33-44. doi: 10.1055/s-2008-1050462.
The automatic perimeter TOPCON SBP-1000 uses light emitting diodes (LED's) for stimulus generation and thus completely differs from the projection type perimeter OCTOPUS 201. In order to get some first idea about the capabilities of the TOPCON SBP-1000 22 eyes of 18 patients with early glaucomatous visual field defects were examined both with the glaucoma program/TOPCON SBP-1000 and program 31/OCTOPUS 201. Field defects were divided into three groups according to the total loss (program DELTA/OCTOPUS 201): group I with a total loss of 0 dB, group II with a total loss of 1-200 dB, group III with a total loss greater than 200 dB. Evaluation and comparison of visual field results is done after discussing the most important technical parameters and fundamental aspects of methods and strategies using the total percentage of "normal" test points, "small relative defects", "deep relative defects" and "absolute defects". We could show that specificity and sensitivity for the detection of earliest field defects are quite equal both in the TOPCON SBP-1000 and in the OCTOPUS 201. Concerning the detection of more pronounced visual field defects the TOPCON-perimeter has a somewhat smaller sensitivity as a result of the different definition of the category "normal"; at the same time there is a tendency for showing up more "absolute defects" instead of "deep relative defects". Technical parameters and strategies are discussed critically taking into consideration the practical experience obtained during this first routine testing.
自动视野计TOPCON SBP - 1000使用发光二极管(LED)来产生刺激,因此与投影式视野计OCTOPUS 201完全不同。为了初步了解TOPCON SBP - 1000的性能,我们使用青光眼程序/TOPCON SBP - 1000和程序31/OCTOPUS 201对18例早期青光眼视野缺损患者的22只眼睛进行了检查。根据总缺损量(OCTOPUS 201的DELTA程序)将视野缺损分为三组:第一组总缺损量为0 dB,第二组总缺损量为1 - 200 dB,第三组总缺损量大于200 dB。在讨论了使用“正常”测试点、“小相对缺损”、“深相对缺损”和“绝对缺损”的总百分比的最重要技术参数以及方法和策略的基本方面之后,对视野结果进行了评估和比较。我们可以表明,在检测最早的视野缺损方面,TOPCON SBP - 1000和OCTOPUS 201的特异性和敏感性相当。关于检测更明显的视野缺损,由于“正常”类别的定义不同,TOPCON视野计的敏感性略低;同时,有一种倾向是出现更多的“绝对缺损”而不是“深相对缺损”。结合首次常规测试中获得的实践经验,对技术参数和策略进行了批判性讨论。