Abstract

Comparative examinations were carried out under controlled conditions in 101 eyes of 66 open-angle glaucoma patients with field defects (Stages I-III according to Aulhorn) with the Octopus and Competer computerized perimeters. The results of the Octopus perimetry were selected as a reference system to evaluate the sensitivity of the Competer in identifying early glaucomatous field defects. The interpretation of qualitative and quantitative discrepancies between the two instruments took into account the different sizes of the test fields. Of the glaucomatous field defects identified with the Octopus perimeter, 84.1% could be recognized with the Competer as well; 4% were questionably pathologic and 11.9% appeared to be false-negative in the Competers. The 11.9% false-negative results of the Competer were all due to the fact that they were scotomas beyond the test eccentricity of this perimeter. In 16.9% of the positive Competer findings the scotomas were outside the Competer test field by more than 50% of their total extent, whereas 83.1% were within the Competer test field by more than 50% of their total extent. In the light of these results it may be concluded that the Competer provides good sensitivity to detect early glaucomatous field defects once these are within 20 degrees eccentricity, while more peripherally situated scotomas will be missed. It would be advisable to enlarge the test field of the Competer to 30 degrees eccentricity. The relation of early glaucomatous field changes to disk morphology revealed that in 42.5% of the eyes definite glaucomatous disk changes were not recognized.

摘要

在可控条件下,对66例有视野缺损的开角型青光眼患者(根据奥 Horn 分级为I - III期)的101只眼睛,使用Octopus和Competer电脑视野计进行了对比检查。选择Octopus视野检查结果作为参考系统,以评估Competer识别早期青光眼性视野缺损的敏感性。在解释两种仪器之间的定性和定量差异时,考虑了测试视野的不同大小。在用Octopus视野计识别出的青光眼性视野缺损中,84.1%也能被Competer识别;4%的结果可疑为病理性,11.9%在Competer视野计中显示为假阴性。Competer的11.9%假阴性结果均是由于这些暗点超出了该视野计的测试偏心度范围。在Competer的阳性结果中,16.9%的暗点超过其总面积的50%位于Competer测试视野之外,而83.1%的暗点超过其总面积的50%位于Competer测试视野之内。根据这些结果可以得出结论,一旦早期青光眼性视野缺损位于偏心度20度以内,Competer对其检测具有良好的敏感性,而更周边的暗点将会漏检。建议将Competer的测试视野扩大到偏心度30度。早期青光眼性视野变化与视盘形态的关系显示,42.5%的眼睛未识别出明确的青光眼性视盘变化。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验