Wasiq Mohammed Abdul, Behura Sushree Smita, Sahoo Swaranjika, Panda Santosh Kumar
Department of Pediatrics, Kalinga Institute of Medical Sciences, KIIT DU, Bhubaneswar, 751024, Odisha, India.
Eur J Pediatr. 2023 Mar;182(3):1083-1087. doi: 10.1007/s00431-022-04777-1. Epub 2022 Dec 27.
Identifying hyperglycaemia during postoperative period is essential for neonates. The objective of the study was to analyse the accuracy and reliability of continuous glucose monitoring (CGM) device for detecting hyperglycaemia during postoperative period in neonates. In this prospective study, hourly glucose recordings by CGM device and six hourly by glucometer glucose (GG-reference test for patient management) were collected in ten surgical neonates during first three postoperative days. Mean absolute relative difference (MARD) and proportion of CGM values within ± 15%/15 mg/dL, ± 20%/20 mg/dL, and ± 30%/30 mg/dL of GG, were analysed from matched pair CGM and GG recordings. The diagnostic performance of CGM for neonatal hyperglycaemia (> 150 mg/dL) was expressed as sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV); and the clinical reliability was assessed in Clarke Error Grid Analysis. A total of 720 CGM glucose readings, 120 matched -paired glucose readings by CGM --GG, and 37 episodes were hyperglycaemia by GG. The MARD was 10.76 %; the percentages of glucose readings within 15%/15 mg/dL, 20%/20 mg/dL, and 30%/30 mg/dL were 94.6%, 97.3% and 100% during the hyperglycaemia period. The sensitivity, specificity, PPV and NPV to detect hyperglycaaemia by CGM device were 100%, 93.9%, 88% and 100 % respectively. In Clarke Error Grid Analysis, 97.3 % points were in zone A and B during the hyperglycaemia period.
CGM device can be a clinically reliable tool for hyperglycaemia management during postoperative period in neonates.
• Neonates are vulnerable for hyperglycaemia during post-operative period and bed side glucometers are used for frequent glucose monitoring in them. • Continuous glucose monitor(CGM) devices are used for the glucose monitoring in adult and paediatric diabetes care.
• For the first time, this study analysed the accuracy and clinical reliability of FreeStyle Libre (CGM device) for identifying hyperglycaemia during post-operative period in neonates. • CGM device has very good accuracy for detecting hyperglycaemia in neonates, it may help the clinician for better glucose stability during post-operative period.
术后期间识别新生儿高血糖至关重要。本研究的目的是分析连续血糖监测(CGM)设备在检测新生儿术后高血糖方面的准确性和可靠性。在这项前瞻性研究中,在10例接受手术的新生儿术后的头三天,收集了CGM设备每小时的血糖记录以及血糖仪每六小时的血糖记录(用于患者管理的参考测试)。从配对的CGM和血糖仪血糖记录中分析平均绝对相对差异(MARD)以及CGM值在血糖仪血糖值±15%/15mg/dL、±20%/20mg/dL和±30%/30mg/dL范围内的比例。CGM对新生儿高血糖(>150mg/dL)的诊断性能以灵敏度、特异性、阳性预测值(PPV)和阴性预测值(NPV)表示;并在克拉克误差网格分析中评估临床可靠性。总共获得了720次CGM血糖读数、120对CGM与血糖仪配对的血糖读数,且血糖仪检测到37次高血糖事件。MARD为10.76%;在高血糖期间,血糖读数在15%/15mg/dL、20%/20mg/dL和30%/30mg/dL范围内的百分比分别为94.6%、97.3%和100%。CGM设备检测高血糖的灵敏度、特异性、PPV和NPV分别为100%、93.9%、88%和100%。在克拉克误差网格分析中,高血糖期间97.3%的点位于A区和B区。
CGM设备可成为新生儿术后高血糖管理的临床可靠工具。
• 新生儿在术后期间易发生高血糖,床边血糖仪用于对其进行频繁血糖监测。• 连续血糖监测(CGM)设备用于成人和儿童糖尿病护理中的血糖监测。
• 本研究首次分析了FreeStyle Libre(CGM设备)在识别新生儿术后高血糖方面的准确性和临床可靠性。• CGM设备在检测新生儿高血糖方面具有很高的准确性,可能有助于临床医生在术后期间更好地维持血糖稳定。