Department of Physical Therapy, Occupational Therapy, Rehabilitation and Physical Medicine, Universidad Rey Juan Carlos, Alcorcón, Spain; Cátedra Institucional en Docencia, Clínica e Investigación en Fisioterapia: Terapia Manual, Punción Seca y Ejercicio Terapéutico, Universidad Rey Juan Carlos, Alcorcón, Madrid, Spain.
Phys Med. 2023 Feb;106:102530. doi: 10.1016/j.ejmp.2023.102530. Epub 2023 Jan 10.
Despite panoramic ultrasound imaging (US) is a promising advance for the morphological and histological assessment of large musculature which cannot be entirely assessed using B-mode, there is no evidence assessing if this technology produces muscle deformation during imaging acquisition. We aimed to analyze differences in size, shape and brightness descriptors between B-mode and panoramic US images and to assess the concordance between both methods. Methods We analyzed size (cross-sectional area and perimeter), shape (circularity, aspect ratio and roundness) and brightness (mean echo-intensity) features of cervical multifidus (CM) and short rotators (SR) in 46 healthy volunteers. Images were acquired in B-mode and extended field-of-view mode. For validity analysis, mean differences between methods were calculated. For agreement analysis, intraclass correlation coefficients (ICCs), standard error of measurements (SEM), minimal detectable changes (MDC) and coefficient of variation (CV%) were calculated.
All parameters showed no significant differences between both methods for either CM or SR (P > 0.05). Panoramic US showed excellent concordance with B-mode for assessing all CM parameters (all ICCs > 0.9), while for SR the agreement ranged from good-to-excellent (ICC from 0.861 to 0.978).
Panoramic US seems to be a valid tool for assessment of muscle size, shape and brightness as no deformation in comparison with B-mode images was seen. Further research is needed to corroborate these findings comparing panoramic US imaging with Gold Standard methods.
尽管全景超声成像(US)是一种很有前途的技术,可以对使用 B 模式无法完全评估的大肌肉进行形态学和组织学评估,但尚无证据表明该技术在成像采集过程中会导致肌肉变形。我们旨在分析 B 模式和全景 US 图像之间在大小、形状和亮度描述符方面的差异,并评估两种方法之间的一致性。
我们分析了 46 名健康志愿者的颈多裂肌(CM)和短旋转肌(SR)的大小(横截面积和周长)、形状(圆形度、纵横比和圆度)和亮度(平均回波强度)特征。图像分别在 B 模式和扩展视场模式下获取。为了进行有效性分析,计算了两种方法之间的平均差异。为了进行一致性分析,计算了组内相关系数(ICCs)、测量标准误差(SEM)、最小可检测变化(MDC)和变异系数(CV%)。
对于 CM 或 SR,两种方法之间的所有参数均无显著差异(P > 0.05)。全景 US 在评估所有 CM 参数时与 B 模式具有极好的一致性(所有 ICCs > 0.9),而对于 SR,一致性范围为良好至极好(ICC 为 0.861 至 0.978)。
全景 US 似乎是一种有效的评估肌肉大小、形状和亮度的工具,与 B 模式图像相比,没有发现变形。需要进一步的研究来证实这些发现,即将全景 US 成像与金标准方法进行比较。