J Zoo Wildl Med. 2023 Jan;53(4):777-784. doi: 10.1638/2022-0027.
Noninvasive blood pressure measurement is commonly performed with oscillometry; however, this technique provides clinically helpful information only if it is representative of the gold standard. Agreement between direct and oscillometric blood pressure measurements were performed in 14 anesthetized, captive tigers (). A cuff, placed around the tail base and connected to a multiparameter monitor, was used to measure arterial blood pressure oscillometrically and provided systolic, mean, and diastolic pressures. At the same time, direct blood pressures were obtained from a dorsal pedal arterial catheter, and the oscillometric and direct readings were considered paired data points. Agreement between the two methods was evaluated by Bland-Altman plots. All animals completed the study and provided 196 paired data points. The bias (mm Hg) for systolic, mean, and diastolic arterial pressures was -3.7, -0.8, and -1.6, respectively. Limits of agreement (mm Hg) for systolic, mean, and diastolic arterial pressures were -31 to 24, -29 to 27, and -29 to 26, respectively. Oscillometry provided an acceptable amount of readings within 10 and 20 mm Hg of the gold standard. The oscillometric technique provided reasonable agreement with direct measurements. Therefore, in the conditions used in this study, oscillometric blood pressure measured via the ventral coccygeal artery provided reasonable estimates of invasive blood pressure in anesthetized tigers.
非侵入性血压测量通常采用振荡法进行;然而,只有当这种技术能够代表金标准时,它才能提供有临床价值的信息。在 14 只麻醉的圈养老虎中进行了直接和振荡血压测量的一致性评估。一个袖带环绕尾巴基部并连接到多参数监测仪上,用于振荡法测量动脉血压,并提供收缩压、平均压和舒张压。同时,从背侧跖动脉导管直接测量血压,将振荡法和直接读数视为配对数据点。通过 Bland-Altman 图评估两种方法之间的一致性。所有动物均完成了研究,并提供了 196 对配对数据点。收缩压、平均压和舒张压的偏倚(mmHg)分别为-3.7、-0.8 和-1.6。收缩压、平均压和舒张压的一致性界限(mmHg)分别为-31 至 24、-29 至 27 和-29 至 26。振荡法在 10 至 20mmHg 的范围内提供了可接受数量的读数。振荡技术与直接测量具有合理的一致性。因此,在本研究中使用的条件下,通过腹侧尾骨动脉测量的振荡血压能够合理估计麻醉老虎的侵入性血压。