Abstract

Uridine contains the chromophore uracil, a base forming part of RNA. In the range 240-290 nm, the absorption spectra of uridine and DNA are very similar and correspond to the spectral inactivation sensitivity of almost all microorganisms. This makes the uridine (absorption maximum 262 nm) an ideal actinometer for determining the germicidal photon flux in the range of 240 to 290 nm. Uridine actinometry is a simple, environmental-friendly, and easy-to-operate actinometry. Thanks to the uridine absorbance spectrum, it was found to be a perfect fit for the photon flux validation of UVC systems. Conventional UV disinfection systems are generally based on low-pressure (LP) mercury lamps which emit at 254 nm. On the other hand, UV light-emitting diodes (UV-LEDs) are a relatively new source of UV light for water treatment, emitting at various wavelengths. This protocol suggests an accurate, simple, easy to operate and straightforward way to determine the photon flux of UVC systems. •Because of the uridine absorbance spectrum, it is an ideal actinometer for photon flux validation of UVC systems.•Initial uridine concentration and photoproduct absorbance impact the kinetic order and quantum yield.•The protocol for UVC uridine actinometry is appropriate for UV-LP and UV-LED sources for water disinfection.

摘要

尿苷含有发色团尿嘧啶,它是构成RNA的一种碱基。在240 - 290纳米范围内,尿苷和DNA的吸收光谱非常相似,且与几乎所有微生物的光谱失活敏感性相对应。这使得尿苷(最大吸收波长为262纳米)成为测定240至290纳米范围内杀菌光子通量的理想光量计。尿苷光量测定法是一种简单、环保且易于操作的光量测定法。由于尿苷的吸收光谱,它被发现非常适合用于UVC系统的光子通量验证。传统的紫外线消毒系统通常基于发射254纳米波长光的低压汞灯。另一方面,紫外线发光二极管(UV-LED)是一种相对较新的用于水处理的紫外线光源,可发射多种波长的光。本方案提出了一种准确、简单、易于操作且直接的方法来测定UVC系统的光子通量。•由于尿苷的吸收光谱,它是用于UVC系统光子通量验证的理想光量计。•初始尿苷浓度和光产物吸光度会影响动力学级数和量子产率。•UVC尿苷光量测定法的方案适用于用于水消毒的紫外线低压汞灯和UV-LED光源。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/efdf/9845997/d775f07e0bec/ga1.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验