日常生活中腕式光电容积脉搏波描记术检测心房颤动的准确性。

Accuracy of wristwatch-type photoplethysmography in detecting atrial fibrillation in daily life.

作者信息

Nonoguchi Noriko Matsushita, Soejima Kyoko, Goda Ayumi, Nishimura Kunihiro, Onozuka Daisuke, Fujita Shin, Koyama Fumio, Takano Yuichi, Iguchi Shiho, Sato Hideki, Mohri Takato, Katusme Yumi, Tashiro Mika, Hoshida Kyoko, Miwa Yosuke, Togashi Ikuko, Ueda Akiko, Sato Toshiaki, Kohno Takashi

机构信息

Division of Cardiovascular Medicine, Kyorin University, 6-20-2 Shinkawa, Mitaka-City, Tokyo 181-8611, Japan.

Statistics and Data Analysis, National Cerebral and Cardiovascular Center Research Institute, 6-1 Kishibe-Shimmachi, Suita, Osaka 564-8565, Japan.

出版信息

Eur Heart J Digit Health. 2022 Aug 16;3(3):455-464. doi: 10.1093/ehjdh/ztac041. eCollection 2022 Sep.

Abstract

AIMS

Detection of asymptomatic paroxysmal atrial fibrillation is challenging. Smartphone- or smartwatch-based photoplethysmography is efficient at detecting irregular rhythms using pulse waves but is too complex for older patients. We aimed to evaluate the detection accuracy of atrial fibrillation by a wristwatch-type continuous pulse wave monitor (PWM) in daily life.

METHODS AND RESULTS

Patients at high risk of atrial fibrillation but with no history of atrial fibrillation ( = 163; mean CHADS score, 1.9) and patients with known atrial fibrillation ( = 123, including 34 with persistent atrial fibrillation) underwent PWM and telemetry electrocardiogram recording for 3 days. Risk of atrial fibrillation was judged using the 'Kyorin Atrial Fibrillation Risk Score', a scoring system based on previously reported atrial fibrillation risk scoring systems. The PWM assessed the presence of atrial fibrillation at 30 min intervals, and the results were compared with the telemetry electrocardiogram findings. The PWMs accurately diagnosed two patients with paroxysmal atrial fibrillation in the high-risk group. The PWMs accurately diagnosed 48 of the 55 patients with atrial fibrillation in the known-atrial fibrillation group. The PWM accuracy in detecting patients with atrial fibrillation was as follows: sensitivity, 98.0%; specificity, 90.6%; positive predictive value, 69.4%; negative predictive value, 99.5%. The respective values for intervals with atrial fibrillation were 86.9%, 98.8%, 89.6%, and 98.5%.

CONCLUSION

The wristwatch-type PWM has shown feasibility in detecting atrial fibrillation in daily life and showed the possibility of being used as a screening tool.

摘要

目的

检测无症状阵发性心房颤动具有挑战性。基于智能手机或智能手表的光电容积脉搏波描记法在使用脉搏波检测不规则心律方面很有效,但对老年患者来说过于复杂。我们旨在评估一款手表式连续脉搏波监测仪(PWM)在日常生活中检测心房颤动的准确性。

方法与结果

有房颤高风险但无房颤病史的患者(n = 163;平均CHADS评分1.9)和已知房颤患者(n = 123,包括34例持续性房颤患者)接受了3天的PWM和遥测心电图记录。使用“杏林房颤风险评分”判断房颤风险,该评分系统基于先前报道的房颤风险评分系统。PWM每隔30分钟评估一次心房颤动的存在情况,并将结果与遥测心电图结果进行比较。PWM准确诊断出高风险组中的两名阵发性心房颤动患者。在已知房颤组的55例房颤患者中,PWM准确诊断出48例。PWM检测房颤患者的准确性如下:敏感性为98.0%;特异性为90.6%;阳性预测值为69.4%;阴性预测值为99.5%。房颤发作间期的相应值分别为86.9%、98.8%、89.6%和98.5%。

结论

手表式PWM已显示出在日常生活中检测心房颤动的可行性,并显示出用作筛查工具的可能性。