Bryant Rodney A, Bundy Matthew F
Fire Research Division, National Institute of Standards and Technology, Gaithersburg, MD 20899, USA.
Fire Technol. 2021;57(3). doi: 10.1007/s10694-020-01066-x.
The accuracy of the exhaust flow measurement contributes significantly to the uncertainty of calorimetry measurements for large fire testing. Less than ideal flow characteristics such as skewed velocity distributions are typical of these large-scale flows and make it difficult to achieve the desired measurement accuracy. Consensus standards for fire testing recommend either bi-directional probes or orifice plates to determine exhaust flow. Both have limited accuracy in the presence of less than ideal flow conditions. Averaging pitot probes are an off-the-shelf technology widely used to monitor flows for industrial processes. They have been utilized in a system of large fire calorimeters to demonstrate differences of less than 5% between heat release rate measurements by oxygen consumption calorimetry and the theoretical heat output from a gas burner. Differences exceeded 5% for a small set of conditions but were still less than 10%. Both levels of agreement are within the confirmation requirements of the consensus standards and were achieved without a system calibration as recommended by the standards. Including this technology as an alternate method to measure exhaust flow would be an improvement to relevant fire testing standards and to the overall accuracy of calorimetry measurements for large fire testing.
对于大型火灾测试,排气流量测量的准确性对量热法测量的不确定性有重大影响。诸如速度分布不均等不理想的流动特性是这些大规模气流的典型特征,这使得难以达到所需的测量精度。火灾测试的共识标准推荐使用双向探头或孔板来确定排气流量。在流动条件不理想的情况下,两者的精度都有限。平均皮托管探头是一种现成的技术,广泛用于监测工业过程中的流量。它们已被应用于大型火灾量热计系统中,以证明通过氧消耗热法测量的热释放率与气体燃烧器的理论热输出之间的差异小于5%。在一小部分条件下,差异超过了5%,但仍小于10%。这两个一致程度都在共识标准的确认要求范围内,并且在未按照标准建议进行系统校准的情况下实现了。将这项技术作为测量排气流量的替代方法纳入,将改进相关的火灾测试标准以及大型火灾测试量热法测量的整体精度。