The malpositioned Greenfield filter: lessons learned.

作者信息

Otchy D P, Elliott B M

机构信息

Vascular Surgery Service, Brooke Army Medical Center, Fort Sam Houston, TX 78234.

出版信息

Am Surg. 1987 Oct;53(10):580-3.

PMID:3674601
Abstract

It is essential that a malpositioned Greenfield filter be recognized as such immediately. Three of the 21 Greenfield filters inserted in the current study were malpositioned. One was placed in the right renal vein, one in the right common iliac vein, and another in the right inferior vena cava in a patient with caval duplication and left ileofemoral thrombosis. A review of the postprocedure radiographs was performed. It was found that two of the three malpositioned filters were at an appropriate vertebral level, and, therefore, this single criteria was inadequate to judge proper filter positioning. A more reliable parameter was the maximal diameter of the open legs of the filter. The mean diameter of the legs in the correctly positioned filters was 29.9 mm compared with 19.0 mm for the malpositioned filters. The authors recommend that the diameter of the filter legs be measured on all postprocedure radiographs. If this diameter is not 30 +/- 4 mm (two standard deviations), then malposition of the filter should be suspected, regardless of the vertebral level at which the filter lies.

摘要

必须立即识别出位置不当的格林菲尔德滤器。在本研究中植入的21个格林菲尔德滤器中有3个位置不当。一个放置在右肾静脉,一个在右髂总静脉,另一个在一名患有下腔静脉重复畸形和左髂股静脉血栓形成的患者的右下腔静脉。对术后X线片进行了复查。发现三个位置不当的滤器中有两个处于适当的椎体水平,因此,仅这一标准不足以判断滤器的正确位置。一个更可靠的参数是滤器开口支腿的最大直径。正确放置的滤器支腿平均直径为29.9mm,而位置不当的滤器支腿平均直径为19.0mm。作者建议在所有术后X线片上测量滤器支腿的直径。如果该直径不是30±4mm(两个标准差),则应怀疑滤器位置不当,无论滤器位于哪个椎体水平。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验