Eddowes M J
Sensors Department, THORN EMI Central Research Laboratories, Hayes, Middlesex, Great Britain.
Biosensors. 1987;3(1):1-15. doi: 10.1016/0265-928x(87)80009-3.
Models describing the binding of antigens to surface immobilised antibodies have been developed. These models are applied to the assessment of the likely analytical utility of direct immunochemical sensors employing immobilised antibody layers. Fundamental limitations to the device response arising from the equilibrium and kinetic characteristics of the binding reaction and from mass transport constraints imply that measurement over the micromolar to nanomolar concentration range may be practical.
已开发出描述抗原与表面固定抗体结合的模型。这些模型用于评估采用固定化抗体层的直接免疫化学传感器可能的分析效用。结合反应的平衡和动力学特性以及传质限制所导致的设备响应的基本局限性表明,在微摩尔至纳摩尔浓度范围内进行测量可能是可行的。