材料类型、扭矩值和灭菌对扫描体线性位移的影响:一项体外研究。

Effect of material type, torque value, and sterilization on linear displacements of a scan body: An in vitro study.

作者信息

Diker Emre, Terzioglu Hakan, Gouveia Diogo N M, Donmez Mustafa Borga, Seidt Jeremy, Yilmaz Burak

机构信息

Private Practice, Med Dis Dental Clinic, Ankara, Turkey.

Department of Prosthodontics, Faculty of Dentistry, Ankara University, Ankara, Turkey.

出版信息

Clin Implant Dent Relat Res. 2023 Apr;25(2):419-425. doi: 10.1111/cid.13187. Epub 2023 Feb 10.

Abstract

BACKGROUND

There is limited knowledge on the effect of scan body (SB) material type, torque value, and sterilization on linear displacements of implant SBs.

PURPOSE

To evaluate the effect of material type, torque value, and sterilization on linear displacements of SBs during screw tightening by using digital image correlation (DIC) analysis.

MATERIALS AND METHODS

One polyetheretherketone (PEEK, Zfx Intraoral Scan Body) and one titanium SB (Ti, MPS Zimmer Scanbody R1410) were tightened with 5 Ncm torque on two implants (Zimmer TSV ⌀4.7 mm) by using a digital torque limiting device. SBs' initial spatial positions relative to the implants were recorded by using 3D DIC technique. Measurements were repeated after initially increasing torque value to 10 Ncm and then to 15 Ncm, and these steps were repeated for a total of 10 PEEK and 10 Ti SBs on both implants (n = 20). All SBs were then sterilized 25 times by using an autoclave (STATIM 5000 S G4) according to manufacturer's recommendations and all measurements were repeated. Linear displacements on three axes were calculated for each SB with increasing torque values (from 5 to 10 Ncm and from 10 to 15 Ncm) before and after sterilization. SB displacements within each torque value-sterilization pair were compared by using Mann-Whitney U test, whereas Wilcoxon signed-rank test was used to compare SB displacements within each material-torque value pair between conditions and within each material-sterilization pair between torque values (α = 0.05).

RESULTS

On x-axis, PEEK SBs had higher displacements than Ti SBs (p < 0.001), whereas sterilization (p ≤ 0.028) and 15 Ncm torque application (p ≤ 0.006) led to higher displacements of PEEK SBs. On y-axis, PEEK SBs had higher displacements than Ti SBs with 15 Ncm torque application (p ≤ 0.033). A total of 15 Ncm torque-applied PEEK SBs and 10 Ncm torque-applied Ti SBs had higher displacements after sterilization (p ≤ 0.028). Application of 15 Ncm torque led to higher displacements regardless of the material (p ≤ 0.002). On z-axis, PEEK SBs had higher displacements (p ≤ 0.015), except for 10 Ncm torque-applied sterilized SBs (p = 0.102). With 10 Ncm torque application, sterilization decreased the displacement values of PEEK SBs (p = 0.044). Greater displacements were observed with 10 Ncm torque-applied Ti SBs before sterilization and 15 Ncm torque-applied PEEK SBs after sterilization (p ≤ 0.033).

CONCLUSIONS

Axial displacement of SBs was affected by material type, torque value, and sterilization. Ti SBs mostly had lower displacements than PEEK SBs. Application of 15 Ncm torque to tested PEEK SBs should be refrained from and a calibrated tightening tool may enable the application of 10 Ncm or lower torque values for lower displacements. Sterilization generally increased PEEK SB displacements.

摘要

背景

关于扫描体(SB)材料类型、扭矩值和灭菌对种植体SB线性位移的影响,目前了解有限。

目的

通过使用数字图像相关(DIC)分析,评估材料类型、扭矩值和灭菌对拧紧螺钉过程中SB线性位移的影响。

材料与方法

使用数字扭矩限制装置,在两颗种植体(Zimmer TSV ⌀4.7毫米)上,分别以5牛厘米的扭矩拧紧一个聚醚醚酮(PEEK,Zfx口腔内扫描体)扫描体和一个钛制SB(Ti,MPS Zimmer扫描体R1410)。通过3D DIC技术记录SB相对于种植体的初始空间位置。在最初将扭矩值增加到10牛厘米,然后再增加到15牛厘米后重复测量,在两颗种植体上对总共10个PEEK和10个Ti SB重复这些步骤(n = 20)。然后根据制造商的建议,使用高压灭菌器(STATIM 5000 S G4)对所有SB进行25次灭菌,并重复所有测量。计算每个SB在灭菌前后随着扭矩值增加(从5到10牛厘米以及从10到15牛厘米)在三个轴上的线性位移。使用曼-惠特尼U检验比较每个扭矩值-灭菌组合内的SB位移,而使用威尔科克森符号秩检验比较每种材料-扭矩值组合在不同条件下以及每种材料-灭菌组合在不同扭矩值之间的SB位移(α = 0.05)。

结果

在x轴上,PEEK SB的位移高于Ti SB(p < 0.001),而灭菌(p ≤ 0.028)和施加15牛厘米扭矩(p ≤ 0.006)导致PEEK SB的位移更高。在y轴上,在施加15牛厘米扭矩时,PEEK SB的位移高于Ti SB(p ≤ 0.033)。总共施加15牛厘米扭矩的PEEK SB和施加10牛厘米扭矩的Ti SB在灭菌后位移更高(p ≤ 0.028)。无论材料如何,施加15牛厘米扭矩都会导致更高的位移(p ≤ 0.002)。在z轴上,PEEK SB的位移更高(p ≤ 0.015),但施加10牛厘米扭矩的灭菌SB除外(p = 0.102)。在施加10牛厘米扭矩时,灭菌降低了PEEK SB的位移值(p = 0.044)。在灭菌前观察到施加10牛厘米扭矩的Ti SB位移更大,在灭菌后观察到施加15牛厘米扭矩的PEEK SB位移更大(p ≤ 0.033)。

结论

SB的轴向位移受材料类型、扭矩值和灭菌的影响。Ti SB的位移大多低于PEEK SB。应避免对测试的PEEK SB施加15牛厘米的扭矩,并且校准的拧紧工具可能能够施加10牛厘米或更低的扭矩值以实现更小的位移。灭菌通常会增加PEEK SB的位移。