Suppr超能文献

智能手机裂隙灯成像——可用性与质量评估

Smartphone Slit Lamp Imaging-Usability and Quality Assessment.

作者信息

Muth Daniel Rudolf, Blaser Frank, Foa Nastasia, Scherm Pauline, Mayer Wolfgang Johann, Barthelmes Daniel, Zweifel Sandrine Anne

机构信息

Department of Ophthalmology, University Hospital Zurich, University of Zurich, Frauenklinikstrasse 24, 8091 Zurich, Switzerland.

Department of Ophthalmology, University Hospital, LMU Munich, Mathildenstrasse 8, 80336 Munich, Germany.

出版信息

Diagnostics (Basel). 2023 Jan 24;13(3):423. doi: 10.3390/diagnostics13030423.

Abstract

PURPOSE

To assess the usability and image quality of a smartphone adapter for direct slit lamp imaging.

METHODS

A single-center, prospective, clinical study conducted in the Department of Ophthalmology at the University Hospital Zurich, Switzerland. The smartphone group consisted of 26 medical staff (consultants, residents, and students). The control group consisted of one ophthalmic photographer. Both groups took images of the anterior and the posterior eye segment of the same proband. The control group used professional photography equipment. The participant group used an Apple mounted on a slit lamp via a removable smartphone adapter (Custom Surgical GmbH, Munich, Germany). The image quality was graded independently by two blinded ophthalmologists on a scale from 0 (low) to 10 (high quality). Images with a score ≥ 7.0/10 were considered as good as the reference images. The acquisition time was measured. A questionnaire on usability and experience in smartphone and slit lamp use was taken by all of the participants.

RESULTS

Each participant had three attempts at the same task. The overall smartphone quality was 7.2/10 for the anterior and 6.4/10 for the posterior segment. The subjectively perceived difficulty decreased significantly over the course of three attempts (Kendall's W). Image quality increased as well but did not improve significantly from take 1 to take 3. However, the image quality of the posterior segment was significantly, positively correlated (Spearman's Rho) with work experience. The mean acquisition time for anterior segment imaging was faster in the smartphone group compared to the control group (156 vs. 206 s). It was vice versa for the posterior segment (180 vs. 151 s).

CONCLUSION

Slit lamp imaging with the presented smartphone adapter provides high-quality imaging of the anterior segment. Posterior segment imaging remains challenging in terms of image quality. The adapter constitutes a cost-effective, portable, easy-to-use solution for recording ophthalmic photos and videos. It can facilitate clinical documentation and communication among colleagues and with the patient especially outside normal consultation hours. Direct slit lamp imaging allows for time to be saved and increases the independence of ophthalmologists in terms of patient mobility and the availability of photographic staff.

摘要

目的

评估用于直接裂隙灯成像的智能手机适配器的可用性和图像质量。

方法

在瑞士苏黎世大学医院眼科进行的一项单中心、前瞻性临床研究。智能手机组由26名医务人员(顾问、住院医师和学生)组成。对照组由一名眼科摄影师组成。两组均对同一受试者的眼前段和眼后段进行成像。对照组使用专业摄影设备。参与组通过一个可拆卸的智能手机适配器(德国慕尼黑的Custom Surgical GmbH公司)将苹果手机安装在裂隙灯上。由两名不知情的眼科医生独立地将图像质量从0(低)到10(高质量)进行评分。得分≥7.0/10的图像被认为与参考图像质量相当。测量采集时间。所有参与者都填写了一份关于智能手机和裂隙灯使用的可用性及体验的问卷。

结果

每位参与者对同一任务进行了三次尝试。眼前段的智能手机成像总体质量为7.2/10,眼后段为6.4/10。在三次尝试过程中,主观感受到的难度显著降低(肯德尔和谐系数)。图像质量也有所提高,但从第一次到第三次拍摄没有显著改善。然而,眼后段的图像质量与工作经验呈显著正相关(斯皮尔曼等级相关系数)。与对照组相比,智能手机组眼前段成像的平均采集时间更快(156秒对206秒)。眼后段则相反(180秒对151秒)。

结论

使用所介绍的智能手机适配器进行裂隙灯成像可提供高质量的眼前段成像。眼后段成像在图像质量方面仍然具有挑战性。该适配器是一种经济高效、便于携带且易于使用的记录眼科照片和视频的解决方案。它有助于临床记录以及同事之间和与患者的沟通,特别是在正常会诊时间之外。直接裂隙灯成像可节省时间,并提高眼科医生在患者移动性和摄影人员可用性方面的独立性。