Ruzza Alessandro, Ferrari Stefano, Airaldi Matteo, Romano Vito, Ponzin Diego
Fondazione Banca degli Occhi del Veneto, 30174 Venice, Italy.
Department of Medical and Surgical Specialties, Radiological Sciences, and Public Health, Ophthalmology Clinic, University of Brescia, 25121 Brescia, Italy.
J Clin Med. 2023 Jan 28;12(3):1026. doi: 10.3390/jcm12031026.
The aim of the study was to assess different temperature ranges for the preservation of pre-loaded Descemet Membrane Endothelial Keratoplasty (DMEK) grafts in the DMEK RAPID Mini device.
Three groups of 15 DMEK grafts (five per group) were pre-loaded in the DMEK RAPID Mini and preserved in Optisol-GS for 72 h at different temperatures: group A at >8 °C, group B between 2-8 °C and group C at <2 °C. After stripping and preservation, the viability of the endothelium, cell loss and morphology were assessed through light microscopy following trypan blue and alizarin red staining.
Overall mortality was 4.07%, 3.97% and 7.66%, in groups A, B and C, respectively, with percentages of uncovered areas of 0.31%, 1.36% and 0.20% (all > 0.05). Endothelial cell density variation was 5.51%, 3.06% and 2.82% in groups A, B and C, respectively ( = 0.19). Total Endothelial Cell Loss (ECL) was 4.37%, 5.32% and 7.84% in groups A, B and C, respectively ( = 0.39). Endothelial cell morphology was comparable in all three groups.
In the DMEK RAPID Mini, low temperatures (<2 °C) may affect the quality of pre-loaded grafts, inducing a higher ECL after 72 h of preservation, although no significant differences among groups could be proved. Our data would suggest maintaining grafts loaded in the DMEK RAPID Mini at temperatures between 2-8 °C for appropriate preservation.
本研究的目的是评估在DMEK RAPID Mini设备中保存预先装载的Descemet膜内皮角膜移植术(DMEK)移植物的不同温度范围。
将三组15个DMEK移植物(每组5个)预先装载到DMEK RAPID Mini中,并在Optisol-GS中于不同温度下保存72小时:A组温度>8°C,B组温度在2-8°C之间,C组温度<2°C。在剥离和保存后,通过台盼蓝和茜素红染色后的光学显微镜评估内皮细胞的活力、细胞损失和形态。
A组、B组和C组的总体死亡率分别为4.07%、3.97%和7.66%,未覆盖区域的百分比分别为0.31%、1.36%和0.20%(均>0.05)。A组、B组和C组的内皮细胞密度变化分别为5.51%、3.06%和2.82%(P = 0.19)。A组、B组和C组的总内皮细胞损失(ECL)分别为4.37%、5.32%和7.84%(P = 0.39)。三组内皮细胞形态相当。
在DMEK RAPID Mini中,低温(<2°C)可能会影响预先装载的移植物质量,在保存72小时后导致更高的ECL,尽管各组之间未证明有显著差异。我们的数据表明,将装载在DMEK RAPID Mini中的移植物保持在2-8°C之间的温度以进行适当保存。