Bohner Lauren, Hanisch Marcel, Parize Hian, Sesma Newton, Kleinheinz Johannes, Meier Norbert
Department of Cranio-Maxillofacial Surgery, University Hospital Muenster, 48149 Muenster, Germany.
Department of Prosthodontics, School of Dentistry, University of São Paulo, São Paulo 05508-220, Brazil.
J Clin Med. 2023 Jan 31;12(3):1117. doi: 10.3390/jcm12031117.
The purpose of this study was to assess the feasibility of SEMAC + VAT to reduce artifacts induced by dental implant-supported restorations, such as its impact on the image quality. Dental-implant supported restorations were installed in a dry mandible. Magnetic resonance scans were acquired on a 3-Tesla MRI system. Artifact suppression (SEMAC + VAT) was applied with different intensity modes (weak, moderate, strong). Artifacts assessment was performed by measuring the mandible volume increase in MRI images prior (reference dataset) and after installation of dental implant-supported prosthesis. Image quality was assessed by two examiners using a five-point scale. Inter-examiner concordance and correlation analysis was performed with Cronbach's alpha and Spearman's test with a significance level at = 0.05. Mandible volume increased by 60.23% when no artifact suppression method was used. By applying SEMAC + VAT, the volume increase ranged from 17.13% (strong mode) to 32.77% (weak mode). Visualization of mandibular bone was positively correlated with SEMAC intensity degree. SEMAC + VAT reduced MRI artifacts caused by dental-implant supported restorations. A stronger suppression mode improved visualization of mandibular bone in detriment of the scanning time.
本研究的目的是评估SEMAC + VAT减少牙种植体支持修复体引起的伪影的可行性,例如其对图像质量的影响。在干燥的下颌骨中安装牙种植体支持的修复体。在3特斯拉MRI系统上进行磁共振扫描。采用不同强度模式(弱、中、强)应用伪影抑制(SEMAC + VAT)。通过测量牙种植体支持的假体安装之前(参考数据集)和之后MRI图像中的下颌骨体积增加来进行伪影评估。由两名检查人员使用五点量表评估图像质量。使用Cronbach's alpha和Spearman检验进行检查人员间一致性和相关性分析,显著性水平为α = 0.05。未使用伪影抑制方法时,下颌骨体积增加了60.23%。通过应用SEMAC + VAT,体积增加范围从17.13%(强模式)到32.77%(弱模式)。下颌骨可视化与SEMAC强度程度呈正相关。SEMAC + VAT减少了牙种植体支持修复体引起的MRI伪影。更强的抑制模式改善了下颌骨的可视化,但不利于扫描时间。