Odom J V, Maida T M, Dawson W W, Hobson R
Department of Ophthalmology, West Virginia University, Morgantown.
Doc Ophthalmol. 1987 Mar;65(3):297-306. doi: 10.1007/BF00149936.
The pattern electroretinogram has assumed greater clinical and experimental significance because of its inner retinal origins. However, clinical tests may be confounded by an artifact. We tested subjects varying reference electrode position and eye stimulated while employing the Dawson-Trick-Litzkow (DTL) fiber electrode as the active electrode. The presence of a statistically significant artifactual response could not be confirmed. However, the variability of responses elicited with the outer canthus was less and the signal-to-noise ratio greater than with other reference positions.
由于视网膜电图起源于视网膜内层,其已具有更大的临床和实验意义。然而,临床测试可能会受到伪迹的干扰。我们在使用道森-特里克-利茨科夫(DTL)光纤电极作为活性电极时,测试了不同参考电极位置和受刺激眼睛的受试者。无法证实存在具有统计学意义的伪迹反应。然而,与其他参考位置相比,外眦引起的反应变异性较小,信噪比更高。