Dawson W W, Trick G L, Litzkow C A
Invest Ophthalmol Vis Sci. 1979 Sep;18(9):988-91.
Corneal electrodes useful for clinical electroretinography require topical anesthesia, interfere with vision, can abrade, and are not well accepted by most children and many adults. A low mass conductive thread, corneal (DTL) electrode is described and comparatively tested against the Burian-Allen electrode. The DTL electrode was found to have few of the limitations of the hard contact lens electrode. Furthermore, the DTL electrode signal quality was comparable to that of the Burian-Allen electrode and provided less between-patient variability.
用于临床视网膜电图检查的角膜电极需要局部麻醉,会干扰视力,可能造成擦伤,并且大多数儿童和许多成年人都不太接受。本文描述了一种低质量导电丝角膜(DTL)电极,并将其与Burian-Allen电极进行了对比测试。结果发现,DTL电极几乎没有硬接触镜电极的局限性。此外,DTL电极的信号质量与Burian-Allen电极相当,且患者间的变异性较小。