Effectiveness of staff radiation protection devices for interventional cardiology procedures.

机构信息

Institut de radioprotection et de sûreté nucléaire, Pôle santé et environnement, Service de recherche en dosimétrie, Fontenay-aux-Roses, France.

Belgian Nuclear Research Centre, Research in Dosimetric Applications, Mol, Belgium.

出版信息

Phys Med. 2023 Mar;107:102543. doi: 10.1016/j.ejmp.2023.102543. Epub 2023 Feb 11.

Abstract

PURPOSE

To evaluate the effectiveness of currently available radioprotective (RP) devices in reducing the dose to interventional cardiology staff, especially to the eye lens and brain.

METHODS

The performances of five RP devices (masks, caps, patient drapes, staff lead and lead-free aprons and Zero-Gravity (ZG) suspended radiation protection system) were assessed by means of Monte Carlo (MC) simulations. A geometry representative of an interventional cardiology setup was modelled and several configurations, including beam projections and staff distance from the source, were investigated. In addition, measurements on phantoms were performed for masks and drapes.

RESULTS

An average dose reduction of 65% and 25% to the eyes and the brain respectively was obtained for the masks by MC simulations but a strong influence of the design was observed. The cap effectiveness for the brain ranges on average between 13% and 37%. Nevertheless, it was shown that only some upper parts of the brain were protected. There was no significant difference between the effectiveness of lead and lead-free aprons. Of all the devices, the ZG system offered the highest protection to the brain and eye lens and a protection level comparable to the apron for the organs normally covered.

CONCLUSION

All investigated devices showed potential for dose reduction to specific organs. However, for masks, caps and drapes, it strongly depends on the design, exposure conditions and staff position. Therefore, for a clinical use, it is recommended to evaluate their effectiveness in the planned conditions of use.

摘要

目的

评估目前可用的放射防护(RP)设备在降低介入心脏病学工作人员(尤其是眼睛晶状体和大脑)剂量方面的有效性。

方法

通过蒙特卡罗(MC)模拟评估了五种 RP 设备(面罩、帽子、患者覆盖物、工作人员的铅衣和无铅围裙以及零重力(ZG)悬挂辐射防护系统)的性能。模拟了一个具有代表性的介入心脏病学设置的几何形状,并研究了几种配置,包括射线投影和工作人员与源的距离。此外,还对面罩和覆盖物进行了 phantom 测量。

结果

通过 MC 模拟,面罩可使眼睛和大脑的平均剂量分别降低 65%和 25%,但观察到设计的强烈影响。帽子对大脑的有效范围平均在 13%至 37%之间。然而,事实证明只有大脑的某些上部得到了保护。铅衣和无铅围裙的有效性没有显著差异。在所有设备中,ZG 系统为大脑和眼睛晶状体提供了最高的保护,并且对通常覆盖的器官的保护水平与围裙相当。

结论

所有研究的设备都显示出对特定器官减少剂量的潜力。然而,对于面罩、帽子和覆盖物,其强烈取决于设计、暴露条件和工作人员的位置。因此,在临床使用中,建议在计划的使用条件下评估其有效性。