Division of Forensic Toxicology, Armed Forces Medical Examiner System, 115 Purple Heart Drive, Dover AFB, DE 19902, USA.
SNA International, LLC, Contractor Supporting the Armed Forces Medical Examiner System, 500 Montgomery St, Suite 500, Alexandria, VA 22314, USA.
J Anal Toxicol. 2023 May 19;47(5):448-454. doi: 10.1093/jat/bkad012.
Recently, lysergic acid diethylamide (LSD) has become a resurgent drug of abuse. The detection of LSD is problematic because of the low dosage taken by users, light and heat sensitivity of the analyte and the lack of efficient analytical methods. Presented here is the validation of an automated sample preparation method for the analysis of LSD and its primary urinary metabolite, 2-oxo-3-hydroxy-LSD (OHLSD), in urine samples by liquid chromatography-tandem mass spectrometry. Analytes were extracted from urine using an automated Dispersive Pipette XTRaction method on Hamilton STAR and STARlet liquid handling systems. The limit of detection for both analytes was administratively defined at the lowest calibrator used in the experiments, and the limit of quantitation was 0.05 ng/mL for both analytes. All validation criteria were acceptable per Department of Defense Instruction 1010.16 requirements. This method offers an efficient, sensitive analytical solution to routinely evaluate large numbers of urine specimens for LSD in workplace drug deterrence programs.
最近,麦角酸二乙基酰胺(LSD)已成为一种死灰复燃的滥用药物。由于使用者服用的剂量低、分析物对光和热敏感以及缺乏有效的分析方法,因此 LSD 的检测存在问题。本文介绍了一种通过液相色谱-串联质谱法自动分析尿液中 LSD 及其主要代谢物 2-氧-3-羟基-LSD(OHLSD)的自动样品制备方法的验证。使用 Hamilton STAR 和 STARlet 液体处理系统上的自动分散移液管 XTRaction 方法从尿液中提取分析物。两种分析物的检测限均在实验中使用的最低校准剂的行政定义下确定,两种分析物的定量限均为 0.05ng/mL。根据国防部指令 1010.16 的要求,所有验证标准均合格。该方法为在工作场所药物预防计划中常规评估大量尿液样本中的 LSD 提供了一种高效、灵敏的分析解决方案。