Willems J L, Arnaud P, van Bemmel J H, Bourdillon P J, Degani R, Denis B, Graham I, Harms F M, Macfarlane P W, Mazzocca G
University of Leuven, Belgium.
J Am Coll Cardiol. 1987 Dec;10(6):1313-21. doi: 10.1016/s0735-1097(87)80136-5.
In an effort to standardize and evaluate the performance of electrocardiographic computer measurement programs, a 15 lead reference library has been developed based on simultaneously recorded standard 12 lead and orthogonal XYZ lead data. A set of 250 electrocardiograms (ECGs) with selected abnormalities was analyzed by a group of five referee cardiologists and 11 different 12 lead and 6 XYZ computer programs. Attention was focused on the exact determination of the onsets and offsets of P, QRS and T waves. The referees performed their task on highly amplified, selected complexes from the library in a two round process. Median results of the referees coincided best with the median derived from all programs. An analysis of stability proved that the combined program median was a robust reference. However, some individual program results were widely divergent. Paired t tests demonstrated earlier onset for P and QRS (p less than 0.001), as well as later offset for P and T waves in the median 12 lead than in the XYZ results. Significant differences also existed among results obtained by programs analyzing all standard ECG leads at one time, the so-called multilead programs, and those obtained by the conventional standard three lead analysis programs. As a consequence, the derived P, PR, QRS and QT interval measurements varied quite widely among the various programs. Significant differences were also observed among measurements of Q, R and S duration. Some programs showed Q waves that were on the average 6 ms (p less than 0.001) longer than those of others. This may significantly influence diagnostic performance.(ABSTRACT TRUNCATED AT 250 WORDS)
为规范和评估心电图计算机测量程序的性能,基于同步记录的标准12导联和正交XYZ导联数据开发了一个15导联参考库。一组五名心脏病专家裁判和11种不同的12导联及6种XYZ计算机程序对一组250份有选定异常的心电图进行了分析。重点关注P、QRS和T波起始和结束的精确测定。裁判们在两轮过程中对参考库中高度放大的选定复合波执行任务。裁判的中位数结果与所有程序得出的中位数最为吻合。稳定性分析证明,组合程序中位数是一个可靠的参考。然而,一些个别程序的结果差异很大。配对t检验表明,中位数12导联结果中P波和QRS波的起始更早(p<0.001),P波和T波的结束比XYZ结果更晚。同时分析所有标准心电图导联的程序(即所谓的多导联程序)得出的结果与传统标准三导联分析程序得出的结果之间也存在显著差异。因此,不同程序得出的P、PR、QRS和QT间期测量值差异很大。Q、R和S波持续时间的测量结果之间也观察到显著差异。一些程序显示的Q波平均比其他程序长6毫秒(p<0.001)。这可能会显著影响诊断性能。(摘要截断于250字)