J Pers Med. 2023 Jan 30;13(2):254. doi: 10.3390/jpm13020254.
Transcatheter aortic valve implantation (TAVI) is commonly used to treat patients with stenosed aortic valves. CT is one of the crucial steps in planning TAVI to obtain measurements of the aortic annulus to choose an appropriately sized prosthesis. Incorrect measurements can lead to patient-prosthesis mismatch and other complications. However, some patients cannot undergo ECG-gated CT with radiocontrast because of the presence of radiopaque objects in the thorax, arrhythmia, renal failure, etc. Aim: To explore supplementary methods to improve aortic annulus sizing for TAVI by extracardiac measurements.
We included all patients who underwent CT as part of TAVI planning. Measurements of femoral and iliac arteries and the femoral head cross-sectional area were performed.
CT scans of 139 patients were included in this study. Sixty-three patients (45%) were males. Mean age of the female patients was 79.6 ± 7.1 years and of the male patients was 81.3 ± 6.1 years. Mean aortic annulus perimeter among female patients was 74.3 ± 6 mm (range 61.9-88.2) and 83 ± 7.9 mm among male patients (range 70.1-74.3 mm). Mean diameters of common iliac, external iliac, and common femoral arteries were 9.2 ± 1.8, 7.6 ± 1, 7.6 ± 1 mm, respectively, for females and 10.2 ± 1.8, 8.5 ± 1.3, and 8.6 ± 1.4 mm for males. Mean perimeter of the femoral head (average value of right and left femoral heads) among the female patients was 137.8 ± 6.3 mm, and among male patients was 155 ± 9.6 mm. A significant correlation was observed between the perimeter of the aortic annulus and the perimeter of the femoral head (Pearson's R = 0.224). The correlation between the aortic annulus perimeter and the femoral head perimeter was stronger among men than among women (Pearson's R = 0.66 and 0.19, respectively).
Femoral head diameter is associated with annulus size. This may help size the appropriate prosthesis in cases where the measurements by CT are in the border zone if corroborated by clinically driven data.
经导管主动脉瓣植入术(TAVI)常用于治疗主动脉瓣狭窄患者。CT检查是TAVI手术规划中的关键步骤之一,用于测量主动脉瓣环以选择尺寸合适的人工瓣膜。测量不准确可能导致患者与人工瓣膜不匹配及其他并发症。然而,部分患者因胸部存在不透射线物体、心律失常、肾衰竭等原因无法进行心电门控CT增强扫描。目的:探索通过心外测量来改进TAVI主动脉瓣环尺寸测量的辅助方法。
我们纳入了所有作为TAVI手术规划一部分而接受CT检查的患者。对股动脉、髂动脉及股骨头横截面积进行了测量。
本研究纳入了139例患者的CT扫描数据。其中63例(45%)为男性。女性患者的平均年龄为79.6±7.1岁,男性患者为81.3±6.1岁。女性患者的主动脉瓣环平均周长为74.3±6mm(范围61.9 - 88.2mm),男性患者为83±7.9mm(范围70.1 - 74.3mm)。女性患者的髂总动脉、髂外动脉及股总动脉平均直径分别为9.2±1.8mm、7.6±1mm、7.6±1mm,男性患者分别为10.2±1.8mm、8.5±1.3mm、8.6±1.4mm。女性患者股骨头平均周长(左右股骨头平均值)为137.8±6.3mm,男性患者为155±9.6mm。观察到主动脉瓣环周长与股骨头周长之间存在显著相关性(Pearson相关系数R = 0.224)。男性患者中主动脉瓣环周长与股骨头周长的相关性强于女性患者(Pearson相关系数R分别为0.66和0.19)。
股骨头直径与瓣环大小相关。如果经临床数据证实,这可能有助于在CT测量处于临界范围的情况下选择合适尺寸的人工瓣膜。