To assess the sensitivity and specificity of superior visual field tests administered in virtual reality (VR) with eye tracking (VR-ET) and without eye tracking (VR 0 ) for the fulfillment of insurance coverage criteria for functional upper eyelid surgery as compared with standard automated perimetry (SAP).
This prospective cross-sectional study included 78 eyes from 41 patients with ptosis, brow ptosis, and dermatochalasis undergoing functional upper eyelid surgery evaluation. Participants underwent serial superior visual field tests using SAP and VR 0 or VR-ET in randomized order. Fulfillment of insurance coverage criteria for blepharoplasty was defined as a 30% increase in the grid seen from the untaped to the taped state. The main outcome measure was the sensitivity and specificity of VR 0 , VR-ET, and overall VR in meeting insurance coverage criteria as compared with SAP.
VR had a sensitivity of 84.1% and specificity of 67.6%, with no significant difference between VR 0 and VR-ET. SAP agreed on insurance coverage criteria fulfillment with VR 0 in 28 (71.8%) eyes and with VR-ET in 32 (82.1%) eyes. Insurance coverage criteria fulfillment rates varied significantly by diagnosis on SAP ( p = 0.012) but not VR ( p = 0.059).
VR may be an alternative to SAP for functional upper eyelid surgery evaluation. Future studies are needed to determine differences in patient satisfaction, testing and waiting time, and test-retest reliability between VR and SAP.
评估虚拟现实(VR)中带有眼动追踪(VR-ET)和不带眼动追踪(VR0)的优越视野测试在满足功能性上睑手术保险覆盖标准方面的敏感性和特异性,与标准自动视野计(SAP)相比。
本前瞻性横断面研究纳入 41 例因上睑下垂、眉下垂和皮肤松弛而接受功能性上睑手术评估的患者的 78 只眼。参与者接受了 SAP 以及 VR0 或 VR-ET 的序列优越视野测试,测试顺序为随机。从未贴到贴状态的网格增加 30%被定义为满足眼睑成形术保险覆盖标准。主要观察指标为与 SAP 相比,VR0、VR-ET 和整体 VR 满足保险覆盖标准的敏感性和特异性。
VR 的敏感性为 84.1%,特异性为 67.6%,VR0 和 VR-ET 之间无显著差异。SAP 与 VR0 一致,28 只眼(71.8%)符合保险覆盖标准,与 VR-ET 一致,32 只眼(82.1%)符合保险覆盖标准。SAP 上的诊断(p=0.012)而非 VR(p=0.059)的保险覆盖标准符合率差异显著。
VR 可能是功能性上睑手术评估中 SAP 的替代方法。未来需要研究 VR 和 SAP 之间在患者满意度、测试和等待时间以及测试-重测可靠性方面的差异。