Sunnersjö Lotta, Lindström Henrik, Schött Ulf, Törnquist Noa, Kander Thomas
Department of Intensive and Perioperative Care, Skåne University Hospital, 214 28, Malmö, Sweden.
The Medical Faculty, Lund University Sweden. Sölveg, 223 62, Lund, Sweden.
Thromb J. 2023 Mar 2;21(1):23. doi: 10.1186/s12959-023-00468-5.
The use of viscoelastic tests is becoming increasingly popular. There is a paucity of validation of the reproducibility of varying coagulation states. Therefore, we aimed to study the coefficient of variation (CV) for the ROTEM EXTEM parameters clotting time (CT), clot formation time (CFT), alpha-angle and maximum clot firmness (MCF) in blood with varying degrees of coagulation strength. The hypothesis was that CV increases in states of hypocoagulability.
Critically ill patients and patients subjected to neurosurgery at a university hospital during three separate periods were included. Each blood sample was tested in eight parallel channels, yielding the CVs for the tested variables. In 25 patients, the blood samples were analysed both at baseline and after dilution with albumin 5%, as well as after being spiked with fibrinogen, simulating weak and strong coagulation.
In total, 225 unique blood samples were collected from 91 patients. All samples were analysed in eight parallel ROTEM channels, resulting in 1,800 measurements. In hypocoagulable samples, defined as those with values outside the normal reference range, the CV of CT was higher (median (interquartile range)) (6.3% (5.1-9.5)) than for normocoagulable samples (5.1% (3.6-7.5)), p < 0.001. CFT showed no difference (p = 0.14), while the CV of alpha-angle was higher in hypocoagulable samples (3.6% (2.5-4.6)) than in normocoagulable samples (1.1% (0.8-1.6), p < 0.001. The CV of MCF was higher in hypocoagulable samples (1.8% (1.3-2.6)) than in normocoagulable samples (1.2% (0.9-1.7)), p < 0.001. The CV ranges for the different variables were as follows: CT: 1.2%-37%, CFT: 1.7%-30%, alpha-angle: 0.0%-17% and MCF: 0.0%-8.1%.
CVs for the EXTEM ROTEM parameters CT, alpha-angle, and MCF increased in hypocoagulable blood compared to blood with normal coagulation, confirming the hypothesis for CT, alpha-angle, and MCF but not for CFT. Furthermore, the CVs for CT and CFT were much higher than those for alpha-angle and MCF. The results demonstrate that EXTEM ROTEM results from patients with weak coagulation should be interpreted with the notion of limited precision and that procoagulative treatment, based only on ROTEM EXTEM, should be given with some caution.
粘弹性测试的应用越来越广泛。关于不同凝血状态可重复性的验证研究较少。因此,我们旨在研究不同凝血强度血液中ROTEM EXTEM参数凝血时间(CT)、凝血形成时间(CFT)、α角和最大凝血硬度(MCF)的变异系数(CV)。假设是在低凝状态下CV会增加。
纳入一所大学医院在三个不同时期的重症患者和接受神经外科手术的患者。每个血样在八个平行通道中进行测试,得出测试变量的CV。在25名患者中,对血样在基线时、用5%白蛋白稀释后以及加入纤维蛋白原模拟弱凝血和强凝血后进行分析。
总共从91名患者中采集了225份独特的血样。所有样本在八个平行的ROTEM通道中进行分析,共得到1800次测量结果。在定义为值超出正常参考范围的低凝样本中,CT的CV更高(中位数(四分位间距))(6.3%(5.1 - 9.5)),高于正常凝血样本(5.1%(3.6 - 7.5)),p < 0.001。CFT无差异(p = 0.14),而低凝样本中α角的CV(3.6%(2.5 - 4.6))高于正常凝血样本(1.1%(0.8 - 1.6)),p < 0.001。低凝样本中MCF的CV(1.8%(1.3 - 2.6))高于正常凝血样本(1.2%(0.9 - 1.7)),p < 0.001。不同变量的CV范围如下:CT:1.2% - 37%,CFT:1.7% - 30%,α角:0.0% - 17%,MCF:0.0% - 8.1%。
与正常凝血血液相比,EXTEM ROTEM参数CT、α角和MCF在低凝血液中的CV增加,证实了关于CT、α角和MCF的假设,但CFT未得到证实。此外,CT和CFT的CV远高于α角和MCF的CV。结果表明,对于凝血功能较弱患者的EXTEM ROTEM结果应在精度有限的概念下进行解读,并且仅基于ROTEM EXTEM进行促凝治疗时应谨慎。