Hudson Darren, Abrantes Luis
MRI Clinical Lead, InHealth Group, High Wycombe, United Kingdom.
Head of Operations, South, InHealth Group, High Wycombe, United Kingdom.
BJR Case Rep. 2022 Oct 5;9(1):20220012. doi: 10.1259/bjrcr.20220012. eCollection 2023 Feb.
A patient attended for an MRI scan wearing Heattech thermal clothing. Following the scan, the patient experienced a heating and sunburn sensation over their back. Further investigation has highlighted one similar event to this internationally due to the clothing technology used. The intention of this report is to raise awareness of the potential for this clothing technology to cause thermal injury when worn within MRI, as well as to further emphasise the importance of assessing patient clothing prior to scanning.
一名患者穿着发热衣进行了磁共振成像(MRI)扫描。扫描后,患者背部出现发热和晒伤的感觉。进一步调查发现,由于所使用的服装技术,在国际上有一起与此类似的事件。本报告的目的是提高人们对这种服装技术在MRI环境中穿着时可能导致热损伤的认识,并进一步强调在扫描前评估患者衣物的重要性。