California Medical Innovations Institute, San Diego, CA, USA.
Chongqing Engineering Research Center of Medical Electronics and Information Technology, Chongqing University of Posts and Telecommunications, Chongqing, China.
Dig Dis Sci. 2023 Jun;68(6):2501-2507. doi: 10.1007/s10620-023-07889-2. Epub 2023 Mar 7.
Only limited data exist on repeatability of anorectal studies with the established physiological and clinical technologies for assessment of anorectal function. Fecobionics is a new multi-sensor simulated feces that provide data by integrating elements from current tests.
To study repeatability of anorectal data obtained with the Fecobionics device.
We assessed the database of Fecobionics studies to determine how many repeated studies were done. From a total of 260 Fecobionics studies, 19 subjects with repeated studies using approximately the same protocol and prototype were identified. Key pressure and bending parameters were assessed and the repeatability analyzed using Bland Altman plots. Furthermore, the inter- and intra-individual coefficient of variation (CV) were computed.
Fifteen subjects (5F/10 M) with repeated studies were normal subjects, three were patients with fecal incontinence and one subject suffered from chronic constipation. The main analysis was conducted on the cohort of normal subjects. The bias for 11 parameters were within the confidence interval, whereas two were slightly outside. The interindividual CV was lowest for the bend angle (10.1-10.7) and between 16.3 and 51.6 for the pressure parameters. The intra-individual CVs were approximately half of the inter-individual CVs, spanning from 9.7 to 27.6.
All data from normal subjects were within previously defined normality. The Fecobionics data showed acceptable repeatability with bias within the confidence limits for almost all parameters. The intra-individual CV was much lower than the inter-individual CV. Dedicated large-scale studies are warranted to evaluate the influence of age, sex, and disease on repeatability as well as comparing between technologies.
目前用于评估肛肠功能的既定生理和临床技术,对肛肠研究的可重复性仅有有限的数据。Fecobionics 是一种新的多传感器模拟粪便,它通过整合来自当前测试的元素来提供数据。
研究使用 Fecobionics 设备获得的肛肠数据的可重复性。
我们评估了 Fecobionics 研究的数据库,以确定进行了多少次重复研究。在总共 260 项 Fecobionics 研究中,确定了 19 名使用大致相同方案和原型进行重复研究的受试者。评估了关键压力和弯曲参数,并使用 Bland Altman 图分析了可重复性。此外,计算了个体内和个体间的变异系数(CV)。
15 名(5 女/10 男)重复研究的受试者为正常受试者,3 名为粪便失禁患者,1 名为慢性便秘患者。主要分析针对正常受试者队列进行。11 个参数的偏差在置信区间内,而有两个参数略超出。弯曲角度的个体间 CV 最低(10.1-10.7),压力参数的个体间 CV 在 16.3 和 51.6 之间。个体内 CV 大约是个体间 CV 的一半,范围从 9.7 到 27.6。
所有正常受试者的数据均在先前定义的正常范围内。Fecobionics 数据的可重复性可接受,几乎所有参数的偏差均在置信限内。个体内 CV 远低于个体间 CV。需要进行专门的大规模研究,以评估年龄、性别和疾病对可重复性的影响,以及比较不同技术之间的差异。