将 Fecobionics 转化为一种能够满足客观会阴下降测量临床需求的技术。

Translating Fecobionics Into a Technique That Addresses Clinical Needs for Objective Perineal Descent Measurements.

机构信息

Department of Surgery, the Chinese University of Hong Kong, Shatin, Hong Kong.

Department of Imaging and Interventional Radiology, the Chinese University of Hong Kong, Shatin, Hong Kong.

出版信息

Clin Transl Gastroenterol. 2021 Apr;12(4):e00342. doi: 10.14309/ctg.0000000000000342.

Abstract

INTRODUCTION

Perineal descent is a phenomenon associated with anorectal dysfunction. It is diagnosed by defecography but subjected to manual measurements on the images/videos and interobserver bias. Fecobionics is a simulated feces for assessing important physiological parameters during defecation. Here, we translate Fecobionics into a new method for estimation of perineal descent based on electronic signals from the embedded inertial measurement units (IMUs).

METHODS

A displacement measurement method by a combined zero-velocity update and gravity compensation algorithm from IMUs was developed. The method was verified in a robot model, which mimicked perineal descent motion.

RESULTS

The method correlated well with the reference (R = 0.9789) and had a deviation from the peak displacement (range 0.25-2.5 cm) of -0.04 ± 0.498 cm. The method was further validated in 5 human experiments with comparison to the benchmark defecography technology (R = 0.79).

DISCUSSION

The proposed technology is objective, i.e., electronic measurements rather than by fluoroscopy or MRI. The development may impact clinical practice by providing a resource-saving and objective technology for diagnosing perineal descent in the many patients suffering from anorectal disorders. The technology may also be used in colon experiments with Fecobionics and for other gastrointestinal devices containing IMUs such as ingestible capsules like the Smartpill.

摘要

简介

会阴下降是一种与肛肠功能障碍相关的现象。它通过排粪造影诊断,但在图像/视频上进行手动测量会存在观察者间偏差。Fecobionics 是一种模拟粪便的方法,用于评估排粪过程中的重要生理参数。在这里,我们将 Fecobionics 转化为一种基于嵌入式惯性测量单元(IMU)电子信号的会阴下降估计新方法。

方法

开发了一种基于 IMU 的零速度更新和重力补偿算法的位移测量方法。该方法在模拟会阴下降运动的机器人模型中进行了验证。

结果

该方法与参考值相关性良好(R = 0.9789),与峰值位移的偏差为-0.04 ± 0.498 cm(范围 0.25-2.5 cm)。该方法在 5 个人体实验中与基准排粪造影技术(R = 0.79)进行了比较,得到了进一步验证。

讨论

所提出的技术是客观的,即电子测量而不是透视或 MRI。该技术的开发可能会对临床实践产生影响,为诊断患有肛肠疾病的许多患者的会阴下降提供一种节省资源且客观的技术。该技术还可用于含有 IMU 的结肠实验,如 Smartpill 等可摄入的胶囊等其他胃肠道设备。