一种用于测量运动员比目鱼肌皮肤温度的材料的红外热成像设备的内间和间内可靠性和可重复性。

Intra- and inter-session reliability and repeatability of an infrared thermography device designed for materials to measure skin temperature of the triceps surae muscle tissue of athletes.

机构信息

Facultad de Enfermería, Fisioterapia y Podología, Universidad Complutense de Madrid, Madrid, Spain.

Departamento de Psicología, Universidad de Valladolid, Valladolid, Valladolid, Spain.

出版信息

PeerJ. 2023 Mar 9;11:e15011. doi: 10.7717/peerj.15011. eCollection 2023.

Abstract

BACKGROUND

Infrared thermography devices have been commonly applied to measure superficial temperature in structural composites and walls. These tools were cheaper than other thermographic devices used to measure superficial human muscle tissue temperature. In addition, infrared thermography has been previously used to assess skin temperature related to muscle tissue conditions in the triceps surae of athletes. Nevertheless, the reliability and repeatability of an infrared thermography device designed for materials, such as the Manual Infrared Camera PCE-TC 30, have yet to be determined to measure skin temperature of the triceps surae muscle tissue of athletes.

OBJECTIVE

The purpose was to determine the procedure's intra- and inter-session reliability and repeatability to determine skin temperature within the Manual Infrared Camera PCE-TC 30 thermography device in the triceps surae muscle tissue of athletes, which was initially designed to measure the superficial temperature of materials.

METHODS

A total of 34 triceps surae muscles were bilaterally assessed from 17 healthy athletes using the Manual Infrared Camera PCE-TC 30 thermography device to determine intra- (at the same day separated by 1 h) and inter-session (at alternate days separated by 48 h) reliability and repeatability of the skin temperature of the soleus, medial and lateral gastrocnemius muscles. The triceps surae complex weas measured by a region of interest of 1 cm through five infrared thermography images for each muscle. Statistical analyses comprised intraclass correlation coefficient (ICC), standard error of measurement (SEM), minimum detectable change (MCD), systematic error of measurement, correlation (), and Bland-Altman plots completed with linear regression models ( ).

RESULTS

Intra- and inter-session measurements of the proposed infrared thermography procedure showed excellent reliability (ICC = 0.968-0.977), measurement errors (SEM = 0.186-0.232 °C; MDC = 0.515-0.643 °C), correlations ( = 0.885-0.953), and did not present significant systematic error of measurements ( > 0.05). Adequate agreement between each pair of measurement moments was presented by the Bland-Altman plots according to the limits of agreement and non-significant linear regression models ( = 0.000-0.019; > 0.05).

CONCLUSIONS

The proposed procedure to determine skin temperature within the Manual Infrared Camera PCE-TC 30 thermography device presented excellent intra- and inter-session reliability and repeatability in athletes' triceps surae muscle tissue. Future studies should consider the SEM and MDC of this procedure to measure the skin temperature of soleus, medial, and lateral gastrocnemius muscles to promote triceps surae muscle prevention and recovery in athletes.

摘要

背景

红外热像仪已广泛应用于测量结构复合材料和墙壁的表面温度。这些工具比用于测量浅层人体肌肉组织温度的其他热成像设备更便宜。此外,红外热像已经被用于评估与运动员比目鱼肌状况相关的皮肤温度。然而,专为材料设计的红外热像仪(例如 Manual Infrared Camera PCE-TC 30)的可靠性和可重复性尚未确定,以测量运动员比目鱼肌组织的皮肤温度。

目的

本研究旨在确定 Manual Infrared Camera PCE-TC 30 热像仪设备内确定运动员比目鱼肌组织皮肤温度的程序的内(同一天间隔 1 小时)和间(隔天间隔 48 小时)可靠性和可重复性,该设备最初设计用于测量材料的表面温度。

方法

使用 Manual Infrared Camera PCE-TC 30 热像仪对 17 名健康运动员的 34 对双侧比目鱼肌进行了评估,以确定单次(同一天间隔 1 小时)和间(隔天间隔 48 小时)测量的可靠性和可重复性,用于测量比目鱼肌、内侧和外侧腓肠肌的皮肤温度。比目鱼肌复合体通过五个红外热像图中的 1 厘米感兴趣区域进行测量。统计分析包括组内相关系数(ICC)、测量误差标准(SEM)、最小可检测变化(MCD)、测量系统误差、相关性()和 Bland-Altman 图与线性回归模型()。

结果

所提出的红外热像程序的内和间测量显示出极好的可靠性(ICC=0.968-0.977)、测量误差(SEM=0.186-0.232°C;MDC=0.515-0.643°C)、相关性(=0.885-0.953),并且没有出现显著的测量系统误差(>0.05)。根据一致性界限和无显著线性回归模型(=0.000-0.019;>0.05),Bland-Altman 图显示了每对测量时刻之间的充分一致性。

结论

在所提出的确定 Manual Infrared Camera PCE-TC 30 热像仪内皮肤温度的程序中,运动员比目鱼肌组织的内和间测量具有极好的可靠性和可重复性。未来的研究应考虑该程序的 SEM 和 MCD,以测量比目鱼肌、内侧和外侧腓肠肌的皮肤温度,以促进运动员比目鱼肌的预防和恢复。