Suppr超能文献

Release of corrosion products by F-75 cobalt base alloy in the rat. III: Effects of a carbon surface coating.

作者信息

Black J, Oppenheimer P, Morris D M, Peduto A M, Clark C C

机构信息

Department of Orthopaedic Surgery, University of Pennsylvania, Philadelphia 19104.

出版信息

J Biomed Mater Res. 1987 Oct;21(10):1213-30. doi: 10.1002/jbm.820211005.

Abstract

A pyrolytic carbon coating was applied to F-75 chromium-cobalt-molybdenum alloy in an effort to reduce the release of corrosion products in vivo. After intramuscular implantation in the rat, a complex pattern of serum and urine concentration elevations of chromium, cobalt, and nickel was seen. The carbon-coated implants released more chromium and cobalt than uncoated controls, as seen by significantly elevated metal concentrations in serum and urine. Animals receiving carbon-coated implants showed a high rate of recurrent implant site inflammation. Neoplastic infiltration of 24 animals with coated implants, but not in any of the 16 animals which received either uncoated F-75 microsphere or poly(ethylene) particulate implants.