用于Paprosky IIIB型骨缺损的锥形带槽与圆柱形股骨柄之间的微动和应力屏蔽

Micromotion and stress shielding between taper fluted and cylindrical femoral stems for Paprosky bone deficiency type IIIB.

作者信息

Limmahakhun S, Adulkasem N, Rungsiyakull C, Jingjit W, Rojanasthien S

机构信息

Department of Orthopaedic Surgery, Faculty of Medicine, Chiang Mai University, 110 Inthawarorot Rd., Suthep, Muang District, Chiang Mai, 50200, Thailand.

Department of Mechanical Engineering, Faculty of Engineer, Chiang Mai University, Chiang Mai, Thailand.

出版信息

Musculoskelet Surg. 2023 Dec;107(4):391-396. doi: 10.1007/s12306-023-00781-2. Epub 2023 Mar 21.

Abstract

PURPOSE

Taper fluted and cylindrical femoral stems have been commonly used for revision surgery with femoral bone loss. Shape and material differences between taper fluted and cylindrical confused surgeons whether superiority and inferiority in comparisons especially for type IIIB Paprosky bone deficiency. This study aims to investigate the micromotion between implant and bone interface and stress distribution around the implant in human cadaveric experiments.

METHODS

Paired comparisons of 7 cadavers between taper fluted stem and cylindrical stem were tested under dynamic cyclic loading.

RESULTS

The results showed appropriate bone-implant micromotions for bone osteointegration could be observed between both taper fluted and cylindrical stems (85 ± 10 μm and 113 ± 80 μm respectively, p = 0.59). Taper fluted stem attained the load distribution from the proximal to distal surrounding bone (from 4.92 ± 2.87 MPa to 2.14 ± 1.43 MPa, p = 0.43), while the cylindrical stem type showed the proximal bone stress shielding (from 2.56 ± 0.76 MPa to 5.23 ± 0.77 MPa, p = 0.03).

CONCLUSION

Both taper fluted and cylindrical femoral stems provided an adequate initial stability with a proper micromotions for bone-implant osteointegration for type IIIB Paprosky femoral bone loss. The taper fluted femoral stem had better biomechanical advantage in terms of greater stress transfer to periprosthetic bone.

摘要

目的

锥形带槽和圆柱形股骨柄常用于伴有股骨骨量丢失的翻修手术。锥形带槽和圆柱形之间的形状和材料差异使外科医生对两者比较时的优劣感到困惑,尤其是对于ⅢB型Paprosky骨缺损。本研究旨在通过人体尸体实验研究植入物与骨界面之间的微动以及植入物周围的应力分布。

方法

对7具尸体的锥形带槽柄和圆柱形柄进行配对比较,在动态循环载荷下进行测试。

结果

结果显示,锥形带槽柄和圆柱形柄之间均可观察到适合骨骨整合的骨-植入物微动(分别为85±10μm和113±80μm,p=0.59)。锥形带槽柄实现了从近端到远端周围骨的载荷分布(从4.92±2.87MPa到2.14±1.43MPa,p=0.43),而圆柱形柄型则显示出近端骨应力遮挡(从2.56±0.76MPa到5.23±0.77MPa,p=0.03)。

结论

对于ⅢB型Paprosky股骨骨量丢失,锥形带槽和圆柱形股骨柄均能提供足够的初始稳定性,并伴有适当的骨-植入物骨整合微动。锥形带槽股骨柄在向假体周围骨传递更大应力方面具有更好的生物力学优势。